Zimmer Tourniquet Cuff 30" (Blue), PLC Dual Port, Single Bladder, Item Number: 60-7070-105, UDI...
Stryker Sustainability Solutions began a recall of Zimmer Tourniquet Cuff 30" (Blue), PLC Dual Port, Single Bladder, Item Number: 60-7070-105, UDI: 00885825013868, Reprocessed Tourniquet Cuff - Product Usage... on . FDA classified it as Class II and gives the reason as "Tourniquet cuffs are potentially mislabeled. If used may lead to (1) Device misuse during procedure (2) Inadequate Limb Compression Force". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1802-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Sustainability Solutions
- Product
- Zimmer Tourniquet Cuff 30" (Blue), PLC Dual Port, Single Bladder, Item Number: 60-7070-105, UDI: 00885825013868, Reprocessed Tourniquet Cuff - Product Usage: Reprocessed tourniquet cuffs are indicated for use in patients who require surgery of the extremities with an expected duration of less than 90 minutes when temporary exsanguination of a limb is desired.
- Product code
- KCY Tourniquet, Pneumatic
- Reason for recall
- Tourniquet cuffs are potentially mislabeled. If used may lead to (1) Device misuse during procedure (2) Inadequate Limb Compression Force.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Beginning 10/09/2019, the firm provided an "URGENT MEDICAL DEVICE RECALL" notification letter and attached Recall Effectiveness Check Form to the affected customers through the sales representatives. The notification letter instructed the customer to do the following: (1) Please check inventory for the devices from the affected lot. (2) Complete the Urgent Medical Device Recall Effectiveness Check Form (page 3) and indicate if any of the affected product remains in inventory. Return the completed form through one of the following methods: (a) email; (b) sales representative; or (c) via fax to 480-763-5345. The form must be completed and returned even if no affected product is found. If any of the affected products have been forwarded to additional facilities (including but not limited to distribution warehouses), please contact these facilities and communicate this recall to them. Please instruct these facilities to complete the Recall Effectiveness Check Form (page 3) and return the completed form either via email email, sales representative, or via fax to 480-763-5345. If the form indicates that any affected product remains in your inventory, a prepaid shipping label will be issued for the return of the product. Credit will be issued to the facility for all devices that are returned prior to January 1, 2020.
- Quantity in commerce
- 90
- Distribution
- US Nationwide distribution including the states of SC, TX, UT, NY, AR
- Product codes and lots
- Lot Number 10237165
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1802-2020 | Class II | Terminated | Voluntary: Firm initiated |