Z-1679-2020Class II

ONCOR Avante Garde #5863472, ONCOR Expression #7360717, ONCOR Impression #5857920, ONCOR Impression Plus...

Siemens Medical Solutions USA, Inc. / initiated 2020-02-18 / Terminated

Siemens Medical Solutions USA, Inc began a recall of ONCOR Avante Garde #5863472, ONCOR Expression #7360717, ONCOR Impression #5857920, ONCOR Impression Plus #5857912 with Digital LINAC Systems Control Console... on . FDA classified it as Class II and gives the reason as "Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1679-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ONCOR Avante Garde #5863472, ONCOR Expression #7360717, ONCOR Impression #5857920, ONCOR Impression Plus #5857912 with Digital LINAC Systems Control Console Software Versions 6, 7, 9, 13
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy.
FDA root cause
Use error
Action
A Customer Safety Advisory Notice was sent to all affected consignees on 2/18/2020 via Adobe Mega Sign or USPS certified mail. The notice describes the interlock scenarios, provides user instructions, and warns the user of potential risks. The interlock scenarios are as follows: MONITOR 2: Monitor 2 total dose coincidence. Intent: This interlock is intended to prevent overdose when the primary dose monitory system fails to complete the treatment. TREATMENT TIME: Preset treatment time coincidence with actual treatment time. Intent: This interlock is intended to prevent overdose if preset treatment time is exceeded. MONITOR SYNC (SW): Excessive accumulated dose deviation detected between the primary and the secondary dose monitoring systems. Intent: This interlock is intended to prevent patient mistreatment due to the inaccurate monitoring of delivered dose. When one or more of the above described interlocks occur, it may be assumed that there is an error in the dose monitor system.The Consumer Safety Advisory Notice states that the Operator's Manual Instructions Update document should be added to the product's manual and all users should be informed of the issues.
Quantity in commerce
1012
Distribution
Worldwide distribution. US consignees in the following states: Alabama, California, Delaware, Florida, Georgia, Iowa, Illinois, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, North Carolina, Nebraska, New Jersey, Nevada, New York, Ohio, Oklahoma, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Wisconsin, West Virginia, Wyoming
Product codes and lots
Serial Numbers 3905, 70-4077, 4082, 70-4283, 70-4368, 5177, 3922, 5120, 5170, 5179, 5205, 5226, 5245, 5282, 5300, 5343, 5385, 5391, 5506, 5665, 4078, 70-4143, 70-4235, 70-4296, 70-4344, 5154, 5281, 5417, 3755, 3761, 3802, 3801, 3901, 3981, 4072, 70-4174, 70-4354, 70-4378, 5087, 5088, 5096, 5095, 5098, 5118, 5148, 5396, 5157, 5821

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1679-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1677-2020Class IITerminatedVoluntary: Firm initiated
Z-1678-2020Class IITerminatedVoluntary: Firm initiated
Z-1679-2020Class IITerminatedVoluntary: Firm initiated
Z-1680-2020Class IITerminatedVoluntary: Firm initiated