Z-2083-2020Class II
NelDerm Hydrogel Dressing (Dressing: 6 x 6 ; Pad: 4 x 4 ) Product #: HB66T
Nelson Innovaions, LLC / initiated 2020-02-11 / Open, Classified / root cause: manufacturing process
Nelson Innovaions, Llc began a recall of NelDerm Hydrogel Dressing (Dressing: 6 x 6 ; Pad: 4 x 4 ) Product #: HB66T on . FDA classified it as Class II and gives the reason as "Foreign material near or along the edge of the hydrogel adhesive within the pouch packaging". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2083-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Nelson Innovaions, LLC
- Product
- NelDerm Hydrogel Dressing (Dressing: 6 x 6 ; Pad: 4 x 4 ) Product #: HB66T
- Product code
- NAE Dressing, Wound, Hydrogel Without Drug And/Or Biologic
- Reason for recall
- Foreign material near or along the edge of the hydrogel adhesive within the pouch packaging
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Nelson Innovations, LLC contacted customers on 2/11/20 via phone call to state reason for recall, health risk and request return of affected product and provided call tags for pick-up of product.
- Quantity in commerce
- 100 units
- Distribution
- CA, NC, PR
- Product codes and lots
- Lot #90971
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2083-2020 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY