TorFlex Transseptal Guiding Sheath ; Model numbers: TF8-32-63-45, TF8-32-63-55, TF8-32-63-90, TF8-32-63-135...
Baylis Medical Company Inc. began a recall of TorFlex Transseptal Guiding Sheath ; Model numbers: TF8-32-63-45, TF8-32-63-55, TF8-32-63-90, TF8-32-63-135, TF85-32-63-37, TF85-32-63-45, TF85-32-63-55... on . FDA classified it as Class II and gives the reason as "There is a remote potential for the sterile packaging of lots of the transseptal guiding sheath to be compromised during shipping and handling. In the case that a packaging defect is not identified and the product is...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2069-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baylis Medical Co., Inc.
- Product
- TorFlex Transseptal Guiding Sheath ; Model numbers: TF8-32-63-45, TF8-32-63-55, TF8-32-63-90, TF8-32-63-135, TF85-32-63-37, TF85-32-63-45, TF85-32-63-55, TF85-32-63-90, TF85-32-63-135, TF8-32-81-45, TF8-32-81-55, TF8-32-81-90, TF85-32-81-37, TF85-32-81-45, TF85-32-81-55, and TF85-32-81-90.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- There is a remote potential for the sterile packaging of lots of the transseptal guiding sheath to be compromised during shipping and handling. In the case that a packaging defect is not identified and the product is used in a patient, there may be a remote risk of infection.
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- The firm sent Urgent Medical Device Recall Notification letters to customers via UPS and/or by email on 07-Oct-2019. The notifications were followed through by a tracking number to facilitate confirmation of package delivery. Customers were informed that all Torflex Transseptal Guiding Sheaths sold separately or as part of the NRG Transseptal kit are affected by this action. Customers were instructed to examine their stock, determine whether they had any of the affected lot numbers in stock, and if so, to quarantine affected product. Customers were requested to complete the acknowledgement form and return it to Baylis within 3 business days by email at [email protected] or by faxing the completed form to 906-602-5671 Attn: Quality Department. Customers were informed that once the completed Acknowledgement form was received by Baylis, they would be contacted by a Baylis Medical Representative within 5 business days to manage the disposition of any identified quarantined product. Customers were instructed to address specific questions regarding the recall to Mrs. Ellen Harfield, Director of Quality at 1-800-276-4416.
- Quantity in commerce
- Total of 59067 units (includes both individual units and those sold in NRG RF Transseptal Kits).
- Distribution
- Product distributed throughout the 50 states and in Washington D.C.
- Product codes and lots
- All lots of the following model numbers: TF85 32 63 90, TFFA031116, TF85 32 63 135, TFFA231116, TF8 32 63 90, TFFL211016, TF85 32 63 45, TFFA261016, TF85 32 63 55, TFFB261016, TF85 32 63 45, TFFC031116, TF8 32 63 45, TFFH031116, TF85 32 63 45, TFFA261016, TF85 32 63 45, TFFD031116, TF8 32 63 90, TFFF031116
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2069-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2070-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.