Z-1905-2020Class II

Visualase Cooled Laser Applicator System Tubing Set kits Product numbers: VCLAS 9735559, 3mm tip/1.65mm...

Medtronic Navigation, Inc / initiated 2019-07-08 / Terminated / root cause: manufacturing process

Medtronic Navigation, Inc. began a recall of Visualase Cooled Laser Applicator System Tubing Set kits Product numbers: VCLAS 9735559, 3mm tip/1.65mm catheter VCLAS 9735560, 10mm tip/1.65mm catheter VCLAS... on . FDA classified it as Class II and gives the reason as "The firm received complaints on the saline tubing of Visualase Cooled Laser Applicator System (VCLAS) kit, including incorrect Luer connector assembly, incorrect drip-chamber assembly, and/or excessive adhesive causing...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1905-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Visualase Cooled Laser Applicator System Tubing Set kits Product numbers: VCLAS 9735559, 3mm tip/1.65mm catheter VCLAS 9735560, 10mm tip/1.65mm catheter VCLAS 9735561, 15mm tip/1.85mm catheter
Product code
GEX Powered Laser Surgical Instrument
Reason for recall
The firm received complaints on the saline tubing of Visualase Cooled Laser Applicator System (VCLAS) kit, including incorrect Luer connector assembly, incorrect drip-chamber assembly, and/or excessive adhesive causing tubing occlusion, all of which could result in disruption or prevention of saline flow.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm issued its first customer notification and customer response form on 7/8/2019. On 7/16/2019, a second notification was issued to allow customers to identify recalled product by lot number. On 10/3/2019, the firm issued a third customer notification to include three additional affected lots. The firm requested customers examine inventory for affected lot numbers and quarantine for return to the firm and to complete the attached customer response form.
Quantity in commerce
4654 units
Distribution
US: AL, AZ, AR, CA, CO, CT, DC, FL, GA, IL, IN, KY, LA, ME, MD, MA, MI, MN, MO, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV, WI OUS: France, Germany, Greece, Iceland, Israel, Spain, Sweden, Switzerland, UK
Product codes and lots
All kits with lot numbers from 0211041602 through 0217695790

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1905-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1905-2020Class IITerminatedVoluntary: Firm initiated