Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2002-2020Class II

Central Venous Catheters are indicated for the assessment of hemodynamic status through right atrial, right...

Abbott Medical / initiated 2020-01-02 / Terminated / root cause: manufacturing process

ICU Medical, Inc. began a recall of Central Venous Catheters are indicated for the assessment of hemodynamic status through right atrial, right ventricular, and pulmonary artery and/or wedge... on . FDA classified it as Class II and gives the reason as "Inability for the guidewire to pass through the needles included with the catheter kits". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2002-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Central Venous Catheters are indicated for the assessment of hemodynamic status through right atrial, right ventricular, and pulmonary artery and/or wedge pressure monitoring for patients. Model # 50456-01
Product code
DQE Catheter, Oximeter, Fiber-Optic
Reason for recall
Inability for the guidewire to pass through the needles included with the catheter kits.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm has contracted Stericycle, Inc., Indianapolis, IN for customer notification, response tracking and effectiveness check. On 02 January 2020, Stericycle, Inc. sent URGENT: MEDICAL DEVICE RECALL notification of the market action to the firm's direct consignees of record.
Quantity in commerce
15
Distribution
US Consignees including in the states of AL, CA, FL, GA, IL, ME, MI, MN, MS, NC, NH, NM, NY, OH, TX, UT, VA and WA OUS - Canada.
Product codes and lots
LOT Number/ UDI - 4350491/0 0840619 03011 7

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2002-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1996-2020Class IITerminatedVoluntary: Firm initiated
Z-1997-2020Class IITerminatedVoluntary: Firm initiated
Z-1998-2020Class IITerminatedVoluntary: Firm initiated
Z-1999-2020Class IITerminatedVoluntary: Firm initiated
Z-2000-2020Class IITerminatedVoluntary: Firm initiated
Z-2001-2020Class IITerminatedVoluntary: Firm initiated
Z-2002-2020Class IITerminatedVoluntary: Firm initiated