Fukushima suction tubes; Fukushima Malleable Tapered Suction, 3 FR, 145 MM; REF: TK-1110 -
TeDan Surgical Innovations LLC began a recall of Fukushima suction tubes; Fukushima Malleable Tapered Suction, 3 FR, 145 MM; REF: TK-1110 - Product Usage: TSI Fukushima Suction Tubes are nonpowered, hand-held... on . FDA classified it as Class II and gives the reason as "There was incorrect raw material used in the production of the identified lot". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1845-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Tedan Surgical Innovations, Inc.
- Product
- Fukushima suction tubes; Fukushima Malleable Tapered Suction, 3 FR, 145 MM; REF: TK-1110 - Product Usage: TSI Fukushima Suction Tubes are nonpowered, hand-held reusable manual surgical instruments for general use intended to be used in various general surgical procedures. They are reusable devices that are connected to the hospital suction system to remove debris and blood from the surgical field.
- Product code
- GEA Cannula, Surgical, General & Plastic Surgery
- Reason for recall
- There was incorrect raw material used in the production of the identified lot.
- Root cause tag
- manufacturing process
- FDA root cause
- Mixed-up of materials/components
- Action
- The recalling firm contacted their customer on 7/16/19 by phone. There was a follow up letter, dated 10/2/19 sent to the consignee. The recalling firm instructed their customer to return the recalled product. For questions contact: [email protected] or at 713-726-0886.
- Quantity in commerce
- 2 units
- Distribution
- US Nationwide distribution in the states of NJ.
- Product codes and lots
- Lot: 0619
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1845-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.