Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1892-2020Class II

3.5mm Aresenal Screw Drill Bit , Ref: 330-35-001, Qty. 70, Rx Only, Non-Sterile, UDI: (01)00812926029696 -

Trilliant Surgical LLC / initiated 2019-10-10 / Terminated / root cause: labeling and UDI

Trilliant Surgical, LLC began a recall of 3.5mm Aresenal Screw Drill Bit , Ref: 330-35-001, Qty. 70, Rx Only, Non-Sterile, UDI: (01)00812926029696 - Product Usage: A drill bit facilitates implant... on . FDA classified it as Class II and gives the reason as "Labeling Error; specifically, drill bits from lots # TSL008495A are marked with the lot # of TSL008495. This could lead to parts being misidentified and impact device traceability". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1892-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Trilliant Surgical LLC
Product
3.5mm Aresenal Screw Drill Bit , Ref: 330-35-001, Qty. 70, Rx Only, Non-Sterile, UDI: (01)00812926029696 - Product Usage: A drill bit facilitates implant insertion
Product code
HTW Bit, Drill
Reason for recall
Labeling Error; specifically, drill bits from lots # TSL008495A are marked with the lot # of TSL008495. This could lead to parts being misidentified and impact device traceability.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
On 10/10/2019 the firm sent a "Urgent: Medical Device Recall" via: FedEx overnight to all consignees in receipt of the recalled product. In addition to informing consignees about the recall, they ask consignees to take the following actions: Check inventory stock for the identified device. 2. Locate and remove and/or discontinue use of identified device. 3. Return all products of the affected device to Trilliant Surgical in accordance with the included Return Response. 4. For questions regarding this communication, contact Trilliant Surgical at 800.495.2919. (Monday Friday, 8:00AM 5:00 CST)
Quantity in commerce
5 drill bits
Distribution
US Nationwide distribution in the state of CA. OUS: None
Product codes and lots
Lot # TSL008495A

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1892-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1892-2020Class IITerminatedVoluntary: Firm initiated