EMPOWR Acetabular System, Non-Sterile, Model Numbers 803-15-023 803-15-026 803-15-027 803-15-028 803-15-029...
Encore Medical, LP began a recall of EMPOWR Acetabular System, Non-Sterile, Model Numbers 803-15-023 803-15-026 803-15-027 803-15-028 803-15-029 803-15-030 Product Usage: The EMPOWR Acetabular... on . FDA classified it as Class II and gives the reason as "The firm received complaints that the EMPOWR Acetabular Impactor Balls were fracturing/cracking during impaction which could cause debris to migrate or be left in the joint space". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1929-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Encore Medical, LP
- Product
- EMPOWR Acetabular System, Non-Sterile, Model Numbers 803-15-023 803-15-026 803-15-027 803-15-028 803-15-029 803-15-030 Product Usage: The EMPOWR Acetabular Impactor Balls is intended for use in joint replacement in patients suffering from disability due to noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the natural femoral head; rheumatoid arthritis, correction of functional deformity, femoral fracture, and in the salvage of previously failed surgical attempts.
- Product code
- LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Reason for recall
- The firm received complaints that the EMPOWR Acetabular Impactor Balls were fracturing/cracking during impaction which could cause debris to migrate or be left in the joint space.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm requested its agents to identify and return affected devices.
- Quantity in commerce
- 86 units
- Distribution
- US Nationwide distribution including in the states of MN, VA, IL, IN, SC, N, TX, CA.
- Product codes and lots
- All lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1929-2020 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY