Z-1942-2020Class III
Insufflation tubing to be used with varies insufflator units. REF 031200-10; REF 031322-10; REF 031532-10...
Karl Storz Endoscopy-America, Inc. / initiated 2019-01-09 / Terminated / root cause: supplier
Karl Storz Endoscopy began a recall of Insufflation tubing to be used with varies insufflator units. REF 031200-10; REF 031322-10; REF 031532-10; REF 031622-10 on . FDA classified it as Class III and gives the reason as "The sterile single-use tubing may contain higher ethylene oxide (EO) residues than the acceptable level". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-1942-2020
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Karl Storz Endoscopy-America, Inc.
- Product
- Insufflation tubing to be used with varies insufflator units. REF 031200-10; REF 031322-10; REF 031532-10; REF 031622-10
- Product code
- NKC Tubing/Tubing With Filter, Insufflation, Laparoscopic
- Reason for recall
- The sterile single-use tubing may contain higher ethylene oxide (EO) residues than the acceptable level.
- Root cause tag
- supplier
- FDA root cause
- Vendor change control
- Action
- The recalling firm sent an URGENT SAFETY NOTICE OF PRODUCT RECALL (via FedEx) to the impacted customers on 01/09/20120, along with acknowledgement confirmation for customer to sign and return it to the firm.
- Quantity in commerce
- 543 boxes were distributed; 10 ea per box.
- Distribution
- 57 consignees in the US. AL,AZ,CA,CT,DE,FL,GA,IL,LA,MA,MD,ME,MI,MN,NE,NH,NJ,NY,OH,OK,PA,SC,TX,VA,VT,WA and WI
- Product codes and lots
- Model/Lot Number: 031200-10/Lot W-027781, W-029058, W-028738; 031322-10/ Lot W-027197, W-028171, W-027939, W-028207; 031532-10/ Lot W-026437, W-027332, W-026871; 031622-10/Lot W-029555
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1942-2020 | Class III | Terminated | Voluntary: Firm initiated |