Z-1791-2020Class I

0.035"(.89mm)x80 cm guidewire. Model 114135080. Soft Tip 3.5cm. REF/UDI: 114135080/(01)208863333008067...

Argon Medical Devices, Inc / initiated 2019-12-04 / Terminated / root cause: manufacturing process

Argon Medical Devices, Inc began a recall of 0.035"(.89mm)x80 cm guidewire. Model 114135080. Soft Tip 3.5cm. REF/UDI: 114135080/(01)208863333008067. RxOnly STERILE EO on . FDA classified it as Class I and gives the reason as "The stiffness of the soft end of the guidewire caused tissue perforation". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1791-2020
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Argon Medical Devices, Inc
Product
0.035"(.89mm)x80 cm guidewire. Model 114135080. Soft Tip 3.5cm. REF/UDI: 114135080/(01)208863333008067. RxOnly STERILE EO
Product code
DQX Wire, Guide, Catheter
Reason for recall
The stiffness of the soft end of the guidewire caused tissue perforation.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
On 12/04/2019, the firm sent Medical Device Recall letter by email to their customer/ distributors. Customers are instructed to inspect and quarantine and return any affected product remaining at their facility utilizing the instruction on the Recall Response Form. Customers are instructed to complete the response acknowledgement form included with the letter and return by fax, mail or email as soon as possible. Customers were also instructed to pass the notice on to all those who need to be aware within their organization or to any organization to which the potentially affected devices have been transferred.
Quantity in commerce
1690 units
Distribution
No US consignees. OUS - Sweden, Norway, Denmark, Greece, Slovenia, Turkey, Hong Kong, England
Product codes and lots
Guidewire Model 114135080; LOT numbers: 11235895, 11234476, 11237014, and 11240602.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1791-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1791-2020Class ITerminatedVoluntary: Firm initiated
Z-1792-2020Class ITerminatedVoluntary: Firm initiated