Genicon EZee Single Use Specimen Retrieval Bag -
Genicon, Inc. began a recall of Genicon EZee Single Use Specimen Retrieval Bag - Product Usage: This device is disposable and is packaged and sterilized for single use only. The bag may be... on . FDA classified it as Class II and gives the reason as "Genicon is providing updated IFU labeling to users to provide clarity on how to use the device for multiple specimen retrieval". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2005-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Genicon, Inc.
- Product
- Genicon EZee Single Use Specimen Retrieval Bag - Product Usage: This device is disposable and is packaged and sterilized for single use only. The bag may be used for multiple specimen retraction.
- Product code
- GCJ Laparoscope, General & Plastic Surgery
- Reason for recall
- Genicon is providing updated IFU labeling to users to provide clarity on how to use the device for multiple specimen retrieval.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- A recall notification letter and updated Instructions for Use (IFU) labeling was sent to consignees on 5/1/2020. The firm is asking that labeling be provided to surgeons for them to read and to familiarize themselves with proper use of the specimen bag for multiple retrievals in the same patient. An Acknowlegement/Receipt Form was provided with response required.
- Quantity in commerce
- 1550
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of IL, IN, MO,TN, and the country of Denmark.
- Product codes and lots
- Device Model 550-000-000; UDI 00877972005230; Lot I8920/2020-08-29
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2005-2020 | Class II | Ongoing | Voluntary: Firm initiated |