Z-2005-2020Class II

Genicon EZee Single Use Specimen Retrieval Bag -

Genicon, Inc. / initiated 2018-06-27 / Open, Classified / root cause: labeling and UDI

Genicon, Inc. began a recall of Genicon EZee Single Use Specimen Retrieval Bag - Product Usage: This device is disposable and is packaged and sterilized for single use only. The bag may be... on . FDA classified it as Class II and gives the reason as "Genicon is providing updated IFU labeling to users to provide clarity on how to use the device for multiple specimen retrieval". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2005-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Genicon, Inc.
Product
Genicon EZee Single Use Specimen Retrieval Bag - Product Usage: This device is disposable and is packaged and sterilized for single use only. The bag may be used for multiple specimen retraction.
Product code
GCJ Laparoscope, General & Plastic Surgery
Reason for recall
Genicon is providing updated IFU labeling to users to provide clarity on how to use the device for multiple specimen retrieval.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
A recall notification letter and updated Instructions for Use (IFU) labeling was sent to consignees on 5/1/2020. The firm is asking that labeling be provided to surgeons for them to read and to familiarize themselves with proper use of the specimen bag for multiple retrievals in the same patient. An Acknowlegement/Receipt Form was provided with response required.
Quantity in commerce
1550
Distribution
Worldwide distribution - US Nationwide distribution including in the states of IL, IN, MO,TN, and the country of Denmark.
Product codes and lots
Device Model 550-000-000; UDI 00877972005230; Lot I8920/2020-08-29

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2005-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2005-2020Class IIOngoingVoluntary: Firm initiated