Z-2058-2020Class I
Heartware HVAD Pump Outflow Graft, Product (REF) Number 1125
HeartWare, Inc. / initiated 2020-04-03 / Open, Classified / root cause: design
Heartware, Inc. began a recall of Heartware HVAD Pump Outflow Graft, Product (REF) Number 1125 on . FDA classified it as Class I and gives the reason as "The Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2058-2020
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- HeartWare, Inc.
- Product
- Heartware HVAD Pump Outflow Graft, Product (REF) Number 1125
- Product code
- DSQ Ventricular (Assist) Bypass
- Reason for recall
- The Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm disseminated a Medical Device Safety Alert by letter beginning on 04/03/2020. The notice explained he potential for damage that can occur during the procedure. Specifically, the Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump. The letter also provided additional steps for assembly and attachment to reduce the risk of damage and tearing during the assembly procedure which was provided in Appendix A.
- Quantity in commerce
- 20147 devices
- Distribution
- Worldwide Distribution
- Product codes and lots
- All lots and serial numbers in distribution
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2055-2020 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2056-2020 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2057-2020 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2058-2020 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2059-2020 | Class I | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
