Z-2116-2020Class II

VACUETTE Blood Transfer Unit. Lot no. A19024NX. Expiration date: 2022-02-07 -

C.A. Greiner & Sohne Gesellschaftmbh / initiated 2020-04-03 / Terminated

C.A. Greiner & Sohne Gesellschaftmbh began a recall of VACUETTE Blood Transfer Unit. Lot no. A19024NX. Expiration date: - Product Usage: The Blood Transfer Unit is used to transfer blood from a syringe... on . FDA classified it as Class II and gives the reason as "While transferring blood the needle may dislodged and remain in the tube stopper which may cause needle stick injury or blood leakage". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2116-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
C.A. Greiner & Sohne Gesellschaftmbh
Product
VACUETTE Blood Transfer Unit. Lot no. A19024NX. Expiration date: 2022-02-07 - Product Usage: The Blood Transfer Unit is used to transfer blood from a syringe into the evacuated blood collection tube.
Product code
JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Reason for recall
While transferring blood the needle may dislodged and remain in the tube stopper which may cause needle stick injury or blood leakage.
FDA root cause
Under Investigation by firm
Action
Greiner BIO-ONE notified customers on 04/03/2020 via "Urgent Product Recall-VACUETTE Blood Transfer Unit" letter. The recall letter identified the affected product and lot numbers. The firm requested the customers to immediately isolate the defective product and complete the product disposition form.
Quantity in commerce
90,400 pieces.
Distribution
*requested 4/15/2020
Product codes and lots
Item no. 450225. UDI no. 39120017577547 and 29120017577540.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2116-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2116-2020Class IITerminatedVoluntary: Firm initiated