Z-2427-2020Class II

BioPince Full Core Biopsy Instrument 16ga x 15cm.,Use with Optional CO-Axial Needle MCXS1615BP. QTY.: 5/CS -

Argon Medical Devices, Inc / initiated 2019-12-02 / Terminated / root cause: labeling and UDI

Argon Medical Devices, Inc began a recall of BioPince Full Core Biopsy Instrument 16ga x 15cm.,Use with Optional CO-Axial Needle MCXS1615BP. QTY.: 5/CS - Product Usage: The disposable, automatic BioPince... on . FDA classified it as Class II and gives the reason as "Incorrect product labeling". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2427-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Argon Medical Devices, Inc
Product
BioPince Full Core Biopsy Instrument 16ga x 15cm.,Use with Optional CO-Axial Needle MCXS1615BP. QTY.: 5/CS - Product Usage: The disposable, automatic BioPince Full Core Biopsy Instrument is used to obtain multiple core samples from soft tissue such as liver, kidney, abdominal masses, etc.
Product code
KNW Instrument, Biopsy
Reason for recall
Incorrect product labeling.
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
On December 2, 2019, the firm sent out a "NOTIFICATION OF RECALL" to customers via registered mail, informing customers of the wrong/incorrect, label on the device. The firm requested to their customers to complete the recall form and return it to the firm with the packaged, unused devices. Customers were instructed to inspect, quarantine and return any affected product remaining at their facility utilizing the instruction on the "Recall Response Form". Customers were also instructed to complete the response acknowledgement form included with the letter and return by fax, mail or email as soon as possible to the firm at address provided and to the attention of : Ms. Amy Clendening-Wheeler, RA Manager, email: [email protected]; or Public Contact Information, Rebecca Ellis, VP, RA/QA, Argon Medical Devices 1-469-731-1410 [email protected] Customers were instructed to pass the notice on to all those who need to be aware within their organization or to any organization to which the potentially affected devices have been transferred. For those products in global distribution, Argon to contact the relevant foreign regulatory bodies and conduct similar notifications of foreign distributors on a country by country basis. Returned goods to be reworked with correct labeling. Rework instruction/procedure has been requested by FDA.
Quantity in commerce
165 units in U.S. and 330 units Internationally
Distribution
Worldwide distribution - US Nationwide including in the states of TX, IL, FL, KS, MD, CA, WA, NY, WI, PA and the countries of United Arab Emirates (Dubai), Spain, Canada.
Product codes and lots
Lot No.: 11277175; Model # 370-1580-01, 16ga x 15cm, for use with Optional Co-axial Needle MCXS1615BP

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2427-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2427-2020Class IITerminatedVoluntary: Firm initiated