Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2065-2020Class II

BD Posiflush SF Saline Flush Syringe (US Distributed Catalog 306553), UDI 50382903065535

Becton Dickinson & Co / initiated 2020-04-16 / Terminated / root cause: packaging

Becton Dickinson & Company began a recall of BD Posiflush SF Saline Flush Syringe (US Distributed Catalog 306553), UDI 50382903065535 on . FDA classified it as Class II and gives the reason as "This product has been confirmed to exhibit holes in the packaging, which impacts package integrity and potentially compromises a sterile field". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2065-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD Posiflush SF Saline Flush Syringe (US Distributed Catalog 306553), UDI 50382903065535
Product code
NGT Saline, Vascular Access Flush
Reason for recall
This product has been confirmed to exhibit holes in the packaging, which impacts package integrity and potentially compromises a sterile field.
Root cause tag
packaging
FDA root cause
Under Investigation by firm
Action
On April 16, 2020, the firm initiated customer notification via Urgent Medical Device Recall letters. Customers were informed that the product had been confirmed to exhibit holes in the packaging, potentially compromising a sterile field. Customers were asked to do the following: 1. Immediately review your inventory for the catalog and lot numbers specified in this notification and discard all product subject to the recall. 2. Share this recall notification with any users of the product, within your facilities, to ensure they are also aware of this recall. 3. Complete and return the attached Customer Response Form, to the contact on the form confirming acknowledgement of the notification and request replacement for all discarded product. 4. Report any adverse health consequences experienced with the use of this product to BD. Actions Taken by BD: 1. Corrective actions have been implemented at the manufacturing site to address this issue. 2. As BD is currently experiencing backorders for this product due to increased demand, they are maximizing production to help reduce the impact to customers. 3. BD will work diligently to prioritize the replacement of product to customers affected by the recall. Please complete the Customer Response Form to assure BD is aware of your replacement needs. If you have questions about this recall, you may reach BD Regional Customer Quality at 888-237-2762 OPT 4, OPT 2; Monday through Friday, between 8AM and 5PM (CT).
Quantity in commerce
16,099,310 units
Distribution
Nationwide domestic distribution. Foreign distribution to Austria, Belgium, Denmark, Finland, France, Germany, Italy, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland, UK, Australia, Canada, Mexico, Malaysia, New Zealand
Product codes and lots
Lots 8353952 9011582 9017875 9024676 9045702 9060999 9079716 9127571 9143529 9156595 9163601

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2065-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2065-2020Class IITerminatedVoluntary: Firm initiated