Z-2052-2020Class II

Conformable Thoracic Stent Graft, Rx only, Sterile EO -

W. L. Gore & Associates, Inc. / initiated 2019-12-17 / Terminated / root cause: labeling and UDI

W L Gore & Associates, Inc. began a recall of Conformable Thoracic Stent Graft, Rx only, Sterile EO - Product Usage: This device is intended for endovascular repair of lesions of the descending thoracic... on . FDA classified it as Class II and gives the reason as "The incorrect label was placed on both the primary and secondary package". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2052-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
W. L. Gore & Associates, Inc.
Product
Conformable Thoracic Stent Graft, Rx only, Sterile EO - Product Usage: This device is intended for endovascular repair of lesions of the descending thoracic aorta.
Product code
MIH System, Endovascular Graft, Aortic Aneurysm Treatment
Reason for recall
The incorrect label was placed on both the primary and secondary package.
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
On 12/17/2019, the firm's field representatives hand delivered an URGENT MEDICAL DEVICE RECALL Notification to customers informing them of a label issue with their stents. Two (2) of the 20 units in their work order are labeled with the incorrect part number and size on both the package and carton labels. Customers are instructed to: return the affected products for replacement and If customers have any questions or concern to contact Field Sales Associates for assistance at Telephone number: 03-6746-2560, FAX number: 03-6746-2561
Quantity in commerce
15 stents
Distribution
International the country of Japan.
Product codes and lots
Part number of the device is TGM343420J Serial numbers: 21225065 21225066 21225067 21225068 21225069 21225070 21225071 21225072 21225073 21225074 21225075 21225076 21225077 21225078 21225079 21225080 21225081 21225082 21225083 21225084

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2052-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2052-2020Class IITerminatedVoluntary: Firm initiated