Inpeco FlexLab Centrifuge Module tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions...
Inpeco S.A. began a recall of Inpeco FlexLab Centrifuge Module tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-5-0.elf and CM_1-7-1.H86 or... on . FDA classified it as Class II and gives the reason as "The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1887-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Inpeco S.p.A.
- Product
- Inpeco FlexLab Centrifuge Module tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-5-0.elf and CM_1-7-1.H86 or xCMd_2-5-1-849fad9.elf
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
- Root cause tag
- software
- FDA root cause
- Software design (manufacturing process)
- Action
- The firm sent an Urgent Medical Device Correction letter to all affected customers on March 30, 2020. The letter instructed the customers to complete and return the Customer Letter Receipt Confirmation and Implementation Check form. The firm explained that their regular service provider will contact them to schedule the firmware version upgrade.
- Quantity in commerce
- 65 units
- Distribution
- Worldwide distribution including the following states: Ohio, New York, Virginia, Pennsylvania, Texas, Delaware, California, Florida, Kentucky, Tennessee, Maryland, Rhode Island, Alabama, Massachusetts, Connecticut, Minnesota, Alabama, Montana, Louisiana, New Jersey, Virginia, New Hampshire, Michigan, North Carolina, Washington, Iowa, New Mexico, Nebraska, and Missouri
- Product codes and lots
- Serial Numbers: FLX.002, FLX.006, FLX.008, FLX.011, FLX.0121, FLX.0127, FLX.0128, FLX.0129, FLX.0132, FLX.0133, FLX.0139, FLX.014, FLX.0147, FLX.0149, FLX.0150, FLX.0153, FLX.0154, FLX.0160, FLX.0162, FLX.0167, FLX.0168, FLX.0172, FLX.0173, FLX.0174, FLX.0179, FLX.0180, FLX.0182, FLX.0191, FLX.0192, FLX.0193, FLX.0194, FLX.0197, FLX.0198, FLX.0201, FLX.0202, FLX.0203, FLX.0206, FLX.022, FLX.032, FLX.035, FLX.036, FLX.037, FLX.039, FLX.043, FLX.046, FLX.050, FLX.056, FLX.057, FLX.060, FLX.061, FLX.062, FLX.073, FLX.082, FLX.083, FLX.085, FLX.086, FLX.087, FLX.093, FLX.096, FLX.097, FLX.098, FLX.100, SHD.001, SHD.002, SHD.003
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1887-2020 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1888-2020 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1889-2020 | Class II | Ongoing | Voluntary: Firm initiated |