Z-1888-2020Class II

Abbott Accelerator a3600 Centrifuge Module tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware...

Inpeco S.p.A. / initiated 2020-04-07 / Open, Classified / root cause: software

Inpeco S.A. began a recall of Abbott Accelerator a3600 Centrifuge Module tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-5-0.elf and... on . FDA classified it as Class II and gives the reason as "The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1888-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Inpeco S.p.A.
Product
Abbott Accelerator a3600 Centrifuge Module tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-5-0.elf and CM_1-7-1.H86 or xCMd_2-5-1-849fad9.elf
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
Root cause tag
software
FDA root cause
Software design (manufacturing process)
Action
The firm sent an Urgent Medical Device Correction letter to all affected customers on March 30, 2020. The letter instructed the customers to complete and return the Customer Letter Receipt Confirmation and Implementation Check form. The firm explained that their regular service provider will contact them to schedule the firmware version upgrade.
Quantity in commerce
23 units
Distribution
Worldwide distribution including the following states: Ohio, New York, Virginia, Pennsylvania, Texas, Delaware, California, Florida, Kentucky, Tennessee, Maryland, Rhode Island, Alabama, Massachusetts, Connecticut, Minnesota, Alabama, Montana, Louisiana, New Jersey, Virginia, New Hampshire, Michigan, North Carolina, Washington, Iowa, New Mexico, Nebraska, and Missouri
Product codes and lots
Serial Numbers: ACP.357, ACP.388, ACP.396, ACP.397, ACP.399, ACP.400, ACP.402, ACP.404, ACP.405, ACP.407, ACP.408, ACP.409, ACP.410, ACP.412, ACP.413, ACP.416, ACP.417, ACP.418, ACP.419, ACP.420, ACP.421, ACP.422, ACP.423

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1888-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1887-2020Class IIOngoingVoluntary: Firm initiated
Z-1888-2020Class IIOngoingVoluntary: Firm initiated
Z-1889-2020Class IIOngoingVoluntary: Firm initiated