Z-2003-2020Class II

Inpeco FlexLab, ARCHITECT cSystem c8000/c16000 Interface Module, Part Numbers: FLX-208-00, FLX-208-10 -

Inpeco S.p.A. / initiated 2020-04-07 / Terminated / root cause: software

Inpeco S.A. began a recall of Inpeco FlexLab, ARCHITECT cSystem c8000/c16000 Interface Module, Part Numbers: FLX-208-00, FLX-208-10 - Product Usage: The FlexLab (aka Accelerator a3600) is a... on . FDA classified it as Class II and gives the reason as "When analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2003-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Inpeco S.p.A.
Product
Inpeco FlexLab, ARCHITECT cSystem c8000/c16000 Interface Module, Part Numbers: FLX-208-00, FLX-208-10 - Product Usage: The FlexLab (aka Accelerator a3600) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.
Product code
CEM Electrode, Ion Specific, Potassium
Reason for recall
When analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples.
Root cause tag
software
FDA root cause
Software design
Action
On April 7, 2020, the firm sent an Urgent Medical Device Correction letter to all impacted customers and instructed them to return the Customer Letter Receipt Confirmation and Implementation Check form within 15 days. The customers were asked to confirm that they had read, understood, and applied all the applicable recommended actions. The impacted Customers were told that they would be contacted by their service representative to schedule the system update.
Quantity in commerce
82
Distribution
Worldwide distribution - US Nationwide including in the states of Texas, West Virginia, South Carolina, Delaware, Louisiana, Illinois, Georgia, Florida, New York, Virginia, Arkansas, Wisconsin, California, Minnesota, Missouri, North Carolina, Utah, Michigan, Ohio, Alabama, Oklahoma, Tennessee, Kansas, Hawaii, Arizona, Washington
Product codes and lots
Serial Numbers: AML.001, FLX.0104, FLX.0105, FLX.0106, FLX.0107, FLX.0108, FLX.0112, FLX.0114, FLX.0115, FLX.0116, FLX.0118, FLX.0119, FLX.0120, FLX.0124, FLX.0125, FLX.0126, FLX.0131, FLX.0133, FLX.0134, FLX.0135, FLX.0136, FLX.0140, FLX.0141, FLX.0142, FLX.0143, FLX.015, FLX.0152, FLX.0155, FLX.0158, FLX.0161, FLX.0163, FLX.0164, FLX.017, FLX.0175, FLX.0178, FLX.018, FLX.0184, FLX.0189, FLX.019, FLX.0196, FLX.020, FLX.021, FLX.023, FLX.026, FLX.027, FLX.030, FLX.032, FLX.033, FLX.038, FLX.040, FLX.044, FLX.045, FLX.047, FLX.051, FLX.052, FLX.053, FLX.054, FLX.055, FLX.058, FLX.059, FLX.063, FLX.064, FLX.065, FLX.066, FLX.067, FLX.068, FLX.071, FLX.072, FLX.074, FLX.075, FLX.076, FLX.077, FLX.079, FLX.080, FLX.081, FLX.091, FLX.092, FLX.095, FLX.099, FLX.101, FLX.102, FLX.103

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2003-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2003-2020Class IITerminatedVoluntary: Firm initiated
Z-2004-2020Class IITerminatedVoluntary: Firm initiated