HeartStart MRx Processor Board PCA Replacement Kits 453563478461, a component of the Philips HeartStart MRx...
Philips North America, LLC began a recall of HeartStart MRx Processor Board PCA Replacement Kits 453563478461, a component of the Philips HeartStart MRx Monitorl/Defibrillator - Product Usage: is intended... on . FDA classified it as Class II and gives the reason as "Damaged ESD bags used for storing components in Processor PCA Replacement Kits". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2064-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- HeartStart MRx Processor Board PCA Replacement Kits 453563478461, a component of the Philips HeartStart MRx Monitorl/Defibrillator - Product Usage: is intended for use in hospital and prehospital settings by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support or defibrillation.
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- Damaged ESD bags used for storing components in Processor PCA Replacement Kits
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Philips Healthcare ssued Medical Device Correction letter dated April 10, 2020 via USP stating reason for correction, health risk and action to take: If the Processor Board PCA Replacement Kit has not yet been installed in a Heart5tart MRx Monitor/Defibrillator, quarantine the kit until Philips can replace it. If the PCA has been installed in a HeartStart MRx Monitor/Defibrillator, continue to monitor the status of the device as recommended in the Instructions for Use, including performing Operational Checks routinely as recommended in the Instructions for Use. Verify your MRx does not show a red X nor is it emitting a periodic audio chirp as this could be an indicator a failure has been detected. When the device is turned on, the device will display INOP messages for any failures detected. If a failure has been detected, take the device out of use and contact your local Philips representative. If an affected Processor Board PCA Replacement Kit 453563478461 is identified, please contact your local Phillips representative. Once the customer has contacted their local Philips representative with their HeartStart MRx Monitor/Defibrillator serial number and also confirmed the affected Processor Board PCA serial number, Philips will provide a replacement Processor Board PCA 453563478461 Replacement Kit free of charge. Further information or support concerning this issue, contact your local Philips representative or call us at 1-800-722-9377.
- Quantity in commerce
- 8 units (2 US and 6 OUS)
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of OK, NY and the countries of Libya, Netherlands, Canada, South Africa, United Kingdom.
- Product codes and lots
- Serial Numbers: CO1-1643IDOTDB CO1-1639100N3T C01-16391D0NB2, CO1-1639100N43 C01-1639100ND0 C01-16391D0N6N, C01-16391D0N4N C01-1&39IDON1J
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2064-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
