Z-2136-2020Class II

ISE Electrolyte Reference (2 X 2L) Reagent, REF A28937, UDI: 15099590202491 -

Siemens Healthcare Diagnostics Inc. / initiated 2020-01-29 / Terminated / root cause: manufacturing process

Beckman Coulter Inc. began a recall of ISE Electrolyte Reference (2 X 2L) Reagent, REF A28937, UDI: 15099590202491 - Product Usage: are intended for the quantitative determination of carbon dioxide... on . FDA classified it as Class II and gives the reason as "Electrolyte reference reagent, used for clinical laboratory testing to determine CO2 in human serum or plasma for diagnosis and treatment of potentially serious disorders associated with changes in the body acid-base...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2136-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
ISE Electrolyte Reference (2 X 2L) Reagent, REF A28937, UDI: 15099590202491 - Product Usage: are intended for the quantitative determination of carbon dioxide in human serum or plasma.
Product code
JFL Ph Rate Measurement, Carbon-Dioxide
Reason for recall
Electrolyte reference reagent, used for clinical laboratory testing to determine CO2 in human serum or plasma for diagnosis and treatment of potentially serious disorders associated with changes in the body acid-base balance, has a low pH and may cause CO2 calibration failures, high CO2 quality control recoveries, and false high patient results which can lead to delay in treatment.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
An Urgent Medical Device Recall letter, dated January 23, 2020 was sent on 01/29/2020 to the affected customers via first class mail and/or email. Customers are advised to: 1) Discontinue use of all ISE Electrolyte Reference Reagent lot M909206 kits. 2) Please review the product Safety Data Sheet (SDS) and discard all ISE Electrolyte Reference Reagent lot M909206 kits according to your laboratorys Standard Operating Procedures and/or according to local regulations. The SDS is available for viewing, printing, and download at https://www.beckmancoulter.com/support/tech-docs. 3) Retrospective review of results is not required. CO2 results are acceptable after a successful calibration and when quality control recoveries are within your laboratorys established QC ranges. 4) Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. So that they are assured customers have received this important communication, they request response within 10 days in one of the following ways: " Electronically, if you received this communication via email. " Manually, complete and return the enclosed Response Form. If customers have any questions regarding this notice, please contact Customer Support Center: " From website: http://www.beckmancoulter.com/customersupport/support " By phone: call 1-800-854-3633 in the United States and Canada. " Outside the United States, contact your local Beckman Coulter Representative. If you need replacement product: " For customers in the United States: o Complete the attached Replacement Order Form and email to [email protected] or fax to (866) 294-7850 OR o Call Client Services at (800) 526-3821 " For customers in Canada: o Complete the attached Replacement Order Form and email to Beckmancoultercanada@be
Quantity in commerce
2738 vials
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MS, NC, NH, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV and the countries of CANADA, JAPAN, LEBANON, SAUDI ARABIA, TAIWAN.
Product codes and lots
Lot Number: M909206

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2136-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2136-2020Class IITerminatedVoluntary: Firm initiated