Z-1806-2020Class II

Philips TraumaDiagnost- a Stationary X-Ray System (Bucky System) Model: 720013 -

Philips North America LLC / initiated 2020-04-07 / Terminated / root cause: design

Philips North America, LLC began a recall of Philips TraumaDiagnost- a Stationary X-Ray System (Bucky System) Model: 720013 - Product Usage: general radiography on . FDA classified it as Class II and gives the reason as "If the shaft breaks due to material fatigue, the trauma arm can fall down and may result in injury". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1806-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips TraumaDiagnost- a Stationary X-Ray System (Bucky System) Model: 720013 - Product Usage: general radiography.
Product code
KPR System, X-Ray, Stationary
Reason for recall
If the shaft breaks due to material fatigue, the trauma arm can fall down and may result in injury.
Root cause tag
design
FDA root cause
Device Design
Action
Philips issued Urgent Field Safety Notice (FCO72000015) sent 4/09/20 via certified mail states reason for recall health risk, action to take: The IfU (Instruction for Use) require the operator to not use the system and call for service, if a mechanical defect or malfunction is suspected. As soon as a deformation is determined, the system may no longer be used and call the service engineer. Should you feel uncertain regarding these instructions, please contact Philips. If you would like any further information or support concerning this issue, please contact your local Philips representative.
Quantity in commerce
346 units (5 US and 341 OUS)
Distribution
Worldwide distribution - US Nationwide distribution including the states of CA, ME, RI, TX, WI and the countries of Canada, Australia, Austria, France, Germany, Greece, Hungary, Ireland, Netherlands, Portugal, Slovenia, United Kingdom.
Product codes and lots
All serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1806-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1806-2020Class IITerminatedVoluntary: Firm initiated