Z-1932-2020Class II

Philips Zenition 50, System code 718096 -

Philips North America LLC / initiated 2020-04-10 / Terminated / root cause: design

Philips North America, LLC began a recall of Philips Zenition 50, System code 718096 - Product Usage: The devices are used for radiological guidance and visualization during diagnostic, interventional and... on . FDA classified it as Class II and gives the reason as "When strain relief is lost at the stand side of the Stand MVS Trolley Cable, over time one or more electrical contacts could break due to pulling, twisting, or dropping of the connector. If this happens, live parts may...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1932-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Zenition 50, System code 718096 - Product Usage: The devices are used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
When strain relief is lost at the stand side of the Stand MVS Trolley Cable, over time one or more electrical contacts could break due to pulling, twisting, or dropping of the connector. If this happens, live parts may be exposed (230 V) and/or the video signal can be lost.
Root cause tag
design
FDA root cause
Device Design
Action
On April 10, 2020, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed of the product issue. Customers were asked to do the following: - Ensure that Cabling Check and Connectors Check are performed daily, as indicated in the User Routine Checks Program (section 7.4 of the Instructions for Use provided with the system). - In case the strain relief is lost and/or damage to the cable/connector is identified; stop using the system and contact your Philips local representative so that the cable can be replaced. - Please ensure that all staff working with the system are informed of the content of this safety notice and place a copy of it with the Instructions for Use. Philips will be replacing the Stand MVS Trolley Cable (459800943703) in all affected systems by a new Stand MVS Trolley Cable that has an improved strain relief design. This action will start by end of April 2020. You will be contacted by a local Philips representative to schedule an appointment to replace the cable. If you need any further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377
Quantity in commerce
25 total
Distribution
Worldwide distribution - US Nationwide distribution including in the states of MD and MI. The countries of Austria, Belgium, France, Germany, Netherlands, Norway, South Africa, Spain, Switzerland, and United Kingdom.
Product codes and lots
Serial Numbers 17 7

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1932-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1932-2020Class IITerminatedVoluntary: Firm initiated
Z-1933-2020Class IITerminatedVoluntary: Firm initiated