Elekta MONACO RTP Sytem, radiation treatment planning software system -
Elekta Inc began a recall of Elekta MONACO RTP Sytem, radiation treatment planning software system - Product Usage: The Monaco system is used to make treatment plans for patients with... on . FDA classified it as Class II and gives the reason as "The Monaco RTP Radiation Treatment Planning System may change the shape and volume of the contour potentially resulting in the device delivering an inaccurate dose". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1825-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Elekta Limited
- Product
- Elekta MONACO RTP Sytem, radiation treatment planning software system - Product Usage: The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- The Monaco RTP Radiation Treatment Planning System may change the shape and volume of the contour potentially resulting in the device delivering an inaccurate dose.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- In April of 2020, Elekta Inc. prepared an Urgent/Important Field Safety Notice identifying this issue, the actions that should be taken by the user/customer in order to prevent risks for patients and/or users, and the actions planned by Elekta to correct the problem. PRODUCT: The Monaco RTP System with versions 5.40 or 5.40.01 installed in or after June of 2018. ACTIONS TO BE TAKEN BY THE CUSTOMER OR USER: Customers/users are instructed to follow their standard clinical practice of reviewing plans, including the review of contours and volumes. DVHs that show unexpected overdose or under-dose should be investigated by the customer/user. The customer/user should post the issued notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed and to advise the appropriate personnel who are working with this product on the content of this letter. Customers/users are also required to either acknowledge receipt of this notification via the Elekta Care Community or complete the provided acknowledgement form and return it to Elekta immediately upon receipt, but no later than within 30 days of receiving the notification letter.
- Quantity in commerce
- 96 devices
- Distribution
- Worldwide distribution - US Nationwide Distribution in the states of IA, NJ, NY, PA, TX, WI and the countries of Australia, Canada, China, Denmark, Germany, Hong Kong, Italy, Netherlands, Sweden, Switzerland, Thailand, Turkey, United Kingdom.
- Product codes and lots
- Software Build: 5.40.00 (UDI: (01)00858164002190(10) 5.40.00), 5.40.01 (UDI: 01)00858164002190(10) 5.40.01)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1825-2020 | Class II | Ongoing | Voluntary: Firm initiated |