Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2545-2020Class II

Centurion DIFFICULT IV KIT containing BD's Vacutainer Plastic Lithium Heparin Tubes (BD Vacutainer-BD Codes...

Centurion Medical Products, LP / initiated 2019-06-01 / Terminated / root cause: design

Centurion Medical Products Corporation began a recall of Centurion DIFFICULT IV KIT containing BD's Vacutainer Plastic Lithium Heparin Tubes (BD Vacutainer-BD Codes 367884 & 367960; Centurion code 45177 (VACUTAINER... on . FDA classified it as Class II and gives the reason as "False elevation of carboxyhemoglobin (COHb) results in blood samples collected using BD Vacutainer Plastic Lithium Heparin tubes with the IL GEM 4000 instrument". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2545-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Centurion Medical Products, LP
Product
Centurion DIFFICULT IV KIT containing BD's Vacutainer Plastic Lithium Heparin Tubes (BD Vacutainer-BD Codes 367884 & 367960; Centurion code 45177 (VACUTAINER GREEN TOP) and 84815 (VACUTAINER 3ML LT. GREEN). Centurion Kit Code: IV8860
Product code
JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Reason for recall
False elevation of carboxyhemoglobin (COHb) results in blood samples collected using BD Vacutainer Plastic Lithium Heparin tubes with the IL GEM 4000 instrument.
Root cause tag
design
FDA root cause
Device Design
Action
Centurion Notification issued to parent company, Medline Industries, via email to Recall Team via their Corporate Headquarters on 6/18/19. Letter states reason for recall, health risk and Required Action: 1. Whether or not you have product in inventory, please complete and return the attached Accountability Record via email - [email protected] or via fax -517-546-3356. 2. Please forward the BD notice to any customer(s) that may receive these kits. 3. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program. Contact BO at 1-888-237-2762, (select Option #3 and then Option #2), between 8AM and 5PM CT Monday through Friday.
Quantity in commerce
140 units
Distribution
US Nationwide distribution including in the state of IL.
Product codes and lots
Lot Number: 2019050790 exp. Date: 10/31/2019

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2545-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2545-2020Class IITerminatedVoluntary: Firm initiated