Z-1923-2020Class II

Philips Brilliance IQon Spectral CT, Computed Tomography X-Ray System; Model # 728332; UDI #...

Philips North America LLC / initiated 2020-04-06 / Terminated / root cause: manufacturing process

Philips North America, LLC began a recall of Philips Brilliance IQon Spectral CT, Computed Tomography X-Ray System; Model # 728332; UDI # (01)00884838059542(21)860067 on . FDA classified it as Class II and gives the reason as "The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1923-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Brilliance IQon Spectral CT, Computed Tomography X-Ray System; Model # 728332; UDI # (01)00884838059542(21)860067
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Philips sent a Medical Device Correction-Urgent Field Safety Notice dated April 6, 2020 to all affected customers. The firm instructed customers to review the information with all members of the staff who need to be aware of the contents of the notice and to maintain a copy with the equipment in the Instructions for Use. Customers were directed to complete and return the form stating that they received, read, and understood the contents of the letter and have taken appropriate actions. URGENT - Field Safety Notice Medical Device Correction letter dated 5/20/20 was sent to customers informing them of two additional affected systems. There is no action required by the customer. Testing and engineering analysis demonstrate that the under-torqued bolts do not pose any immediate safety risk as the mechanical connection between the cooling unit and the rotor continues to meet applicable safety standards. The affected systems may continue to be used according to the IFU. Philips is distributing this FSN to all affected customers/users and will correct the issue via Field Change Order FCO 72800725. Urgent Field Safety Notice Medical Device Correction dated 5/20/20 was distributed to notify additional systems not previously included in the original notification. There is no action required by the customer. Testing and engineering analysis demonstrate that the under-torqued bolts do not pose any immediate safety risk as the mechanical connection between the cooling unit and the rotor continues to meet applicable safety standards. The affected systems may continue to be used according to the IFU. Philips is distributing this FSN to all affected customers/users and will correct the issue via Field Change Order FCO 72800725. If you need any further information or support concerning this issue, please contact your local Philips representative: For North America, contact the Customer Care Solutions Center (1-800-722-9377, follow the prompts).
Quantity in commerce
2
Distribution
Worldwide distribution including one unit distributed to Iowa.
Product codes and lots
Serial #s  860067, 860023

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1923-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1921-2020Class IITerminatedVoluntary: Firm initiated
Z-1922-2020Class IITerminatedVoluntary: Firm initiated
Z-1923-2020Class IITerminatedVoluntary: Firm initiated