Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2111-2020Class II

Giraffe Shuttle. Upgrade Kits with part number 2109672-001 and 2109673-001 GTIN: 00840682110808

GE Healthcare / initiated 2020-04-02 / Terminated / root cause: design

GE Healthcare, LLC began a recall of Giraffe Shuttle. Upgrade Kits with part number 2109672-001 and 2109673-001 GTIN: 00840682110808 on . FDA classified it as Class II and gives the reason as "During Giraffe Shuttle charging, the batteries emit low concentrations of gases (outgassing) producing a strong sulfur odor". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2111-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Giraffe Shuttle. Upgrade Kits with part number 2109672-001 and 2109673-001 GTIN: 00840682110808.
Product code
FMZ Incubator, Neonatal
Reason for recall
During Giraffe Shuttle charging, the batteries emit low concentrations of gases (outgassing) producing a strong sulfur odor.
Root cause tag
design
FDA root cause
Device Design
Action
GE Healthcare notified customers on 04/02/2020 via "Urgent Medical Device Correction" letter. The recall letter identified the affected product and serial numbers. The firm instructed the users to discontinue use of the product if at any time they smell a gas odor. GE Healthcare will correct all affected products at a later date when the correction is available at no cost.
Quantity in commerce
204 units.
Distribution
US Nationwide distribution in the states of AZ, CA, CO, DE, FL, GA, ID, IL, IN, LA, MD, MI, MS, NC, ND, NM, NV, NY, OH, Ok, OR, PA, SC, TN, TX, VA, WA, WI and WV.
Product codes and lots
Serial numbers: GSHW60086, GSHX60001, GSHX60014, GSHY70002-GSHY70195, GSHZ70101-GSHZ70108, MAAP50034, MAAQ50170, MAAR50021, MAAT50013, MAAT60045, MAAU61029, MAAU61040 and MAAV6059.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2111-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2111-2020Class IITerminatedVoluntary: Firm initiated