Z-2115-2020Class II

Accelerator a3600 Automation System used with the following Modules: High Volume Storage(P/N FLX-270...

Inpeco S.p.A. / initiated 2020-04-07 / Terminated / root cause: design

Inpeco S.A. began a recall of Accelerator a3600 Automation System used with the following Modules: High Volume Storage(P/N FLX-270, FLX-271, FLX-276, FLX-282, FLX-283), HSQ Interface Module... on . FDA classified it as Class II and gives the reason as "The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by design". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2115-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Inpeco S.p.A.
Product
Accelerator a3600 Automation System used with the following Modules: High Volume Storage(P/N FLX-270, FLX-271, FLX-276, FLX-282, FLX-283), HSQ Interface Module (P/N FLX-274), Aliquoter Module (P/N FLX-209), Rack Output Module ROM400 (P/N FLX-289), XN-9000 Interface Module (P/N FLX-290), Advia 2120 LAS Interface Module (P/N FLX-219), Vertical Transportation Module (P/N FLX -204), Rack Input Module (P/N FLX-214), AUWi System Interface Module (P/N FLX-286), Track To Rack Interface Module (P/N FLX-295) - Product Usage: is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by design.
Root cause tag
design
FDA root cause
Component design/selection
Action
Urgent: Medical Device Correction notification letters dated 3/30/20 were sent to customers. Action to be taken by the user After module protection removal, wait at least 2 seconds before accessing the impacted module (listed in Table 1). Your service provider will contact you to replace the sensor. Until the service visit please maintain awareness on this notice. Please complete and return the "Customer Letter Receipt Confirmation and Implementation Check form" attached to this letter within 15 days directly to the email address specified in the email communication. Please transfer this notice to whom it might concern. Contact reference person: For any clarification you may need, do not hesitate to contact: Eva Balzarotti - Regulatory Affairs Manager E-mail: [email protected] Phone: (+41) 919118 224
Quantity in commerce
2 systems
Distribution
The products were distributed to the following US States: IL and NY.
Product codes and lots
Serial Numbers: ACP.050 ACP.146

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2115-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2114-2020Class IITerminatedVoluntary: Firm initiated
Z-2115-2020Class IITerminatedVoluntary: Firm initiated