Z-2112-2020Class II
CooperSurgical Dilator/Sound Set, Rx only, Sterile/EO, Made in South Korea, P/N MX21
CooperSurgical, Inc. / initiated 2020-04-09 / Terminated / root cause: packaging
CooperSurgical, Inc. began a recall of CooperSurgical Dilator/Sound Set, Rx only, Sterile/EO, Made in South Korea, P/N MX21 on . FDA classified it as Class II and gives the reason as "The seal of the sterile pouch may be compromised on one lot of the Milex Cervical Dilators". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2112-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CooperSurgical, Inc.
- Product
- CooperSurgical Dilator/Sound Set, Rx only, Sterile/EO, Made in South Korea, P/N MX21
- Product code
- HIH Hysteroscope (And Accessories)
- Reason for recall
- The seal of the sterile pouch may be compromised on one lot of the Milex Cervical Dilators.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- CooperSurgical sent an Urgent Medical Device Recall letter dated April 9, 2020 to all customers that received lot number 257589 of the Milex Cervical Dilator P/N MX21 which could have a compromised finished seal (opposite the chevron seal) of the sterile pouch. Customers were instructed to inspect each package for damage, including the seal area, prior to use (product is acceptable for use if it is visually confirmed that the pouchs seal is intact). Customers should discontinue use of and quarantine products with any packaging irregularities. The firm requested that the customers complete and return the attached Acknowledgement and Receipt Form, whether or not they had any of the affected product on hand. CooperSurgical stated that they will arrange for product returns and replacement.
- Quantity in commerce
- 1740 units/174 boxes
- Distribution
- Distribution to North America including the following states: AZ, CA, CT, DC, DE, FL, GA, IA, IN, MA, MI, MN, MO, NC, NH, NY, OH, OK, PA, SC, TN, TX, UT, VA, WI
- Product codes and lots
- Lot # 257589
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2112-2020 | Class II | Terminated | Voluntary: Firm initiated |