Z-1958-2020Class II

Siemens Cios Alpha VA30-mobile X-Ray system, Material # 11105200 The Cios Alpha and Cios Spin are a mobile...

Siemens Medical Solutions USA, Inc. / initiated 2020-04-20 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of Siemens Cios Alpha VA30-mobile X-Ray system, Material # 11105200 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of... on . FDA classified it as Class II and gives the reason as "Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1958-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Cios Alpha VA30-mobile X-Ray system, Material # 11105200 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Siemens Healthcare issued A Customer Safety Advisory Notice dated 4/20/20 via AX030/20/S. Letter states reason for recall, health risk and action to take: We strongly recommend that you do not disconnect the main unit from the monitor trolley when the system displays the system error message Err 16305 / 80: Confirm this error and repeat your last action until the problem has been corrected by a service engineer. Contact our service organization at 1-800-888-7436. Siemens is currently developing a solution to eliminate the root cause of this problem.
Quantity in commerce
16 units
Distribution
US Nationwide distributions.
Product codes and lots
Serial Numbers: 40057 40047 40048 40107 40108 40111 40120 40126 40134 40135 40032 40104 40075 40102 40124 40070

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1958-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1957-2020Class IITerminatedVoluntary: Firm initiated
Z-1958-2020Class IITerminatedVoluntary: Firm initiated
Z-1959-2020Class IITerminatedVoluntary: Firm initiated