Z-2196-2020Class II

Gambro Cartridge Low Weight-Low Volume Blood Transport System for Hemodialysis Blood Set -

Baxter Healthcare Corporation / initiated 2020-04-17 / Terminated / root cause: manufacturing process

Baxter Healthcare Corporation began a recall of Gambro Cartridge Low Weight-Low Volume Blood Transport System for Hemodialysis Blood Set - Product Usage: is intended for single use in a hemodialysis... on . FDA classified it as Class II and gives the reason as "Potential disconnection of tubing set". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2196-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Gambro Cartridge Low Weight-Low Volume Blood Transport System for Hemodialysis Blood Set - Product Usage: is intended for single use in a hemodialysis treatment using Dialysis Delivery Systems.
Product code
FJK Set, Tubing, Blood, With And Without Anti-Regurgitation Valve
Reason for recall
Potential disconnection of tubing set.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Baxter Healthcare notified customers on about 04/17/2020 via "Urgent Device Correction" letter. Customers were instructed to do the following: 1. Prior to use, thoroughly inspect each connection of the set to check for any detachments in the tubing. Operators may continue to use affected sets if no detachments are observed. Additionally, per the Instructions For Use (IFU), users should observe carefully for leaks during priming and use, and examine the tubing carefully to be certain that all connections are secure, all lines are unobstructed and that there are no kinks or leaks in the tubing. 2. If you find sets with tubing disconnections, please contact Baxter Healthcare Center for Service to arrange for return and credit. Baxter Healthcare Center for Service can be reached at 888-229-0001 between the hours of 7:00 am and 6:00 pm Central Time, Monday through Friday. Please have your Baxter 8-digit ship-to account number, product code, lot number, and quantity of product to be returned ready when calling. 3. If you purchased affected product directly from Baxter, a customer reply form is included in your mailing. Please complete the enclosed Baxter customer reply form and return it to Baxter by faxing it to 224-270-5457 or scanning and e-mailing it to [email protected], even if you do not have any inventory. Returning the customer reply form promptly will confirm your receipt of this notification and prevent you from receiving repeat notices. 4. If you purchased this product from a distributor, please note that the Baxter customer reply form is not applicable. If a reply form is provided by your distributor or wholesaler, please return it to the supplier according to their instructions. 5. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that
Quantity in commerce
4,065
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY. Countries of Australia, Austria, Bahrain, Belgium, Canada, Denmark, Egypt, Finland, France, Germany, Greece, Israel, Italy, Kuwait, Lithuania, Latvia, Montenegro, Morocco, Netherlands, Norway, Poland, Russia, Serbia, Slovenia, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom, Korea, Singapore, Hong Kong
Product codes and lots
Product Code: 114435; UDI: 37332414112401; Lot Numbers: 1000212228, 1000215484, 1000224531

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2196-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2193-2020Class IITerminatedVoluntary: Firm initiated
Z-2194-2020Class IITerminatedVoluntary: Firm initiated
Z-2195-2020Class IITerminatedVoluntary: Firm initiated
Z-2196-2020Class IITerminatedVoluntary: Firm initiated