Medtronic Implantable Insulin Pump System, REF: MMT-2007D, Sterile EO, CE 0459 UDI: 00763000043001
Medtronic Inc. began a recall of Medtronic Implantable Insulin Pump System, REF: MMT-2007D, Sterile EO, CE 0459 UDI: 00763000043001 on . FDA classified it as Class II and gives the reason as "Internal testing results showed endotoxin-related test result did not meet specifications for implantable insulin pump. If pumps have higher than allowed endotoxin levels, patients may have an effect of fever...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2206-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Medtronic Implantable Insulin Pump System, REF: MMT-2007D, Sterile EO, CE 0459 UDI: 00763000043001
- Product code
- LKK Pump, Infusion, Implanted, Programmable
- Reason for recall
- Internal testing results showed endotoxin-related test result did not meet specifications for implantable insulin pump. If pumps have higher than allowed endotoxin levels, patients may have an effect of fever, hypotension, anaphylactic shock, hyperglycemia, or diabetic ketoacidosis (DKA), potentially followed by secondary multi-organ failure (primarily renal and hepatic) and/or death.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Medtronic initiated verbal communication with affected consignees on April 10, 2020 via phone call and provided affected consignees with an "Urgent Field Safety Notice" via email between the dates of 04/24/20 - 4/30/20. . In addition to informing consignees about the recall, the firm asked consignees to take the following actions: 1. Immediately identify and quarantine all unused affected pumps in your inventory. 2. Return all unused affected pumps in your inventory to Medtronic. Your Medtronic Representative can assist you in the return and replacement of this pump as necessary. 3. Forward this notice to all those who need to be aware within your organization and to any organization where the affected pumps may have been transferred. Please maintain a copy of this notice in your records.
- Quantity in commerce
- 24 Pumps
- Distribution
- No US Distribution. OUS: Netherlands, France, and Sweden
- Product codes and lots
- Device not distributed within the US. All Pump lots produced May 21, 2019 to present containing the following serial numbers: 31359, 31360, and 31332
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2206-2020 | Class II | Terminated | Voluntary: Firm initiated |