ARTIS icono systems with Quantification Application SW (QVA/QCA) on a standalone XWP computer, an...
Siemens Medical Solutions USA, Inc began a recall of ARTIS icono systems with Quantification Application SW (QVA/QCA) on a standalone XWP computer, an interventional Fluoroscopic X-Ray System Model Number... on . FDA classified it as Class II and gives the reason as "Using Quantification Application (QVA/QCA) on DSA images may lead to a failure in vessel detection or incorrect quantification of vessel detection and result in an incorrect diagnosis and inappropriate treatment of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2257-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- ARTIS icono systems with Quantification Application SW (QVA/QCA) on a standalone XWP computer, an interventional Fluoroscopic X-Ray System Model Number: 11327600
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Using Quantification Application (QVA/QCA) on DSA images may lead to a failure in vessel detection or incorrect quantification of vessel detection and result in an incorrect diagnosis and inappropriate treatment of the patient
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens issued Urgent Medical Device Correction (AX035/20/S) dated 4/23/20. The letter states reason for recall health risk and action to take: We urgently recommend not to use Quantification Application (QVA/QCA) on DSA images until the software problem has been resolved. If Quantification Application has already been used on DSA images in the past, please verify the results and diagnostic evaluation if applicable. Siemens will correct the error with a software patch via Update Instruction AX034/20/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436.
- Quantity in commerce
- 5 US
- Distribution
- AZ, IL, UT, WI
- Product codes and lots
- Serial Numbers: 180014 180015 180018 180019 180029
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2257-2020 | Class II | Terminated | Voluntary: Firm initiated |