Z-2257-2020Class II

ARTIS icono systems with Quantification Application SW (QVA/QCA) on a standalone XWP computer, an...

Siemens Medical Solutions USA, Inc. / initiated 2020-04-23 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of ARTIS icono systems with Quantification Application SW (QVA/QCA) on a standalone XWP computer, an interventional Fluoroscopic X-Ray System Model Number... on . FDA classified it as Class II and gives the reason as "Using Quantification Application (QVA/QCA) on DSA images may lead to a failure in vessel detection or incorrect quantification of vessel detection and result in an incorrect diagnosis and inappropriate treatment of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2257-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ARTIS icono systems with Quantification Application SW (QVA/QCA) on a standalone XWP computer, an interventional Fluoroscopic X-Ray System Model Number: 11327600
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Using Quantification Application (QVA/QCA) on DSA images may lead to a failure in vessel detection or incorrect quantification of vessel detection and result in an incorrect diagnosis and inappropriate treatment of the patient
Root cause tag
software
FDA root cause
Software design
Action
Siemens issued Urgent Medical Device Correction (AX035/20/S) dated 4/23/20. The letter states reason for recall health risk and action to take: We urgently recommend not to use Quantification Application (QVA/QCA) on DSA images until the software problem has been resolved. If Quantification Application has already been used on DSA images in the past, please verify the results and diagnostic evaluation if applicable. Siemens will correct the error with a software patch via Update Instruction AX034/20/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436.
Quantity in commerce
5 US
Distribution
AZ, IL, UT, WI
Product codes and lots
Serial Numbers: 180014 180015 180018 180019 180029

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2257-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2257-2020Class IITerminatedVoluntary: Firm initiated