Z-2473-2020Class II
Elekta Unity, Image-Guided Radiation Therapy System
Elekta Limited / initiated 2019-12-18 / Terminated / root cause: manufacturing process
Elekta, Inc. began a recall of Elekta Unity, Image-Guided Radiation Therapy System on . FDA classified it as Class II and gives the reason as "There have been reports of excessive heating in the MRL Gradient Connection Module". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2473-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Elekta Unity, Image-Guided Radiation Therapy System
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- There have been reports of excessive heating in the MRL Gradient Connection Module.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Initially, the firm issued an email on 12/18/2019 to their field engineers which said they had attached a "Knowledge Article" which provided instructions on how to inspect the Philips Marlin 1.5T system used in the Elekta Unity MR Linac. On 5/11/2020, the firm issued an Urgent Important Field Safety Notification dated May, 2020, via email to consignees which explained the problem, clinical impact, and said that initial corrective actions have been performed on all impacted sites to ensure the gradient coil connection assembly is correct. It goes onto say that extensive testing of the incorrect connection material has been completed to ensure that continued use of the Elekta Unity system is acceptable, however, it remained that further actions can be taken to reduce the risk.
- Quantity in commerce
- 21 devices
- Distribution
- Distributed in US - PA, NY, and IA. Distributed ROW - Australia, Canada, China, Switzerland, Denmark, Hong Kong, Italy, Netherlands, Sweden, Thailand, Turkey, UK. There was no government/military distribution.
- Product codes and lots
- Elekta Unity Serial numbers 600014, 600016, 600017, 600018, 600019, 600020, 600021, 600022, 600023, 600024, 600025, 600026, 600027, 600028, 600029, 600030, 600031, 600032, 600033, 600034, 600035
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2473-2020 | Class II | Terminated | Voluntary: Firm initiated |