Sterile Disposable Connectors ASY Y CONN 3/8 X 3/8 X 3/8 24 REF/ GTIN for Insert Label:050526000/...
LivaNova USA Inc. began a recall of Sterile Disposable Connectors ASY Y CONN 3/8 X 3/8 X 3/8 24 REF/ GTIN for Insert Label:050526000/ 00803622123320 on . FDA classified it as Class II and gives the reason as "The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2455-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LivaNova USA, Inc.
- Product
- Sterile Disposable Connectors ASY Y CONN 3/8 X 3/8 X 3/8 24 REF/ GTIN for Insert Label:050526000/ 00803622123320
- Product code
- DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass
- Reason for recall
- The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- The firm sent an URGENT MEDICAL DEVICE REMOVAL letter to affected customers on 04/24/2020. The purpose of this medical device correction is to: 1. Advise the customers of the present issue and not to use the affected devices; 2. Providing information to the user on product disposal or return and replacement; 3. Inform affected customers that LivaNova will coordinate replacement for the unused products.
- Quantity in commerce
- 984 units
- Distribution
- US - AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV OUS- Spain (Hospital de Barsuto, Bilbao, Spain)
- Product codes and lots
- LOT 2002800191, 2003400085
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2443-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2444-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2445-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2446-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2447-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2448-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2449-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2450-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2451-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2452-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2453-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2454-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2455-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2456-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2457-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2458-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2459-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2460-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2461-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2462-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2463-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2464-2020 | Class II | Terminated | Voluntary: Firm initiated |