Z-2289-2020Class II

Elekta Unity, UDI 05060191071321, Image-guided radiation therapy system -

Elekta Limited / initiated 2020-04-30 / Terminated / root cause: packaging

Elekta, Inc. began a recall of Elekta Unity, UDI 05060191071321, Image-guided radiation therapy system - Product Usage: is indicated for radiation therapy treatments and stereotactic... on . FDA classified it as Class II and gives the reason as "potential risk that Legionella may be present in the Unity machine room, specifically the heat exchanger condensation collection tray". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2289-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Elekta Unity, UDI 05060191071321, Image-guided radiation therapy system - Product Usage: is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
potential risk that Legionella may be present in the Unity machine room, specifically the heat exchanger condensation collection tray.
Root cause tag
packaging
FDA root cause
Under Investigation by firm
Action
The firm notified the service engineers of the issue and the risk to health (exposure to legionella) by email on 04/30/2020. The firm followed with a letter disseminated on 05/12/2020 to customers and service engineers which discussed the hazard and Elekta's updated preventative maintenance.
Quantity in commerce
32 units
Distribution
Worldwide distribution - US Nationwide distribution including in the states of PA, WI, NY, TN, NJ, IA, TX, and the countries of Australia, Canada, China, Denmark, Germany, Hong Kong, Italy, Netherlands, Sweden, Switzerland, Thailand, Turkey, UK.
Product codes and lots
Serial numbers 600001 - 600043

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2289-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2289-2020Class IITerminatedVoluntary: Firm initiated