Elekta Unity, UDI 05060191071321, Image-guided radiation therapy system -
Elekta, Inc. began a recall of Elekta Unity, UDI 05060191071321, Image-guided radiation therapy system - Product Usage: is indicated for radiation therapy treatments and stereotactic... on . FDA classified it as Class II and gives the reason as "potential risk that Legionella may be present in the Unity machine room, specifically the heat exchanger condensation collection tray". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2289-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Elekta Unity, UDI 05060191071321, Image-guided radiation therapy system - Product Usage: is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- potential risk that Legionella may be present in the Unity machine room, specifically the heat exchanger condensation collection tray.
- Root cause tag
- packaging
- FDA root cause
- Under Investigation by firm
- Action
- The firm notified the service engineers of the issue and the risk to health (exposure to legionella) by email on 04/30/2020. The firm followed with a letter disseminated on 05/12/2020 to customers and service engineers which discussed the hazard and Elekta's updated preventative maintenance.
- Quantity in commerce
- 32 units
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of PA, WI, NY, TN, NJ, IA, TX, and the countries of Australia, Canada, China, Denmark, Germany, Hong Kong, Italy, Netherlands, Sweden, Switzerland, Thailand, Turkey, UK.
- Product codes and lots
- Serial numbers 600001 - 600043
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2289-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.