VITROS Chemistry Product AMYL Slides Reagent, AMYLASE/5 PACK/300 SLDS, Product Code 1202670, UDI #...
Ortho-Clinical Diagnostics began a recall of VITROS Chemistry Product AMYL Slides Reagent, AMYLASE/5 PACK/300 SLDS, Product Code 1202670, UDI # 10758750008513 - Product Usage: For in vitro diagnostic use... on . FDA classified it as Class II and gives the reason as "Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2364-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS Chemistry Product AMYL Slides Reagent, AMYLASE/5 PACK/300 SLDS, Product Code 1202670, UDI # 10758750008513 - Product Usage: For in vitro diagnostic use only to quantitatively measure amylase activity in serum, plasma, and urine using VITROS 250/350/5, 1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.
- Product code
- CIW Starch-Dye Bound Polymer, Amylase
- Reason for recall
- Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets.
- FDA root cause
- Under Investigation by firm
- Action
- On May 11, 2020, Ortho Clinical Diagnostics, Inc. sent an Urgent Medical Device Correction Notification to their customers alerting them of important information regarding the precision of VITROS AMYL Slides. The notice stated that the imprecision affects serum, urine and QC samples, however VITROS AMYL Slides performance continues to meet its intended clinical use. The VITROS Performance Verifiers I and II Control Assay Sheet estimated within-lab SD, Range of Means (ROM) and Delta table values were re-evaluated. The Delta tables will be updated in a future revision of the Performance Verifier Training Module. In the meantime, customers were instructed to use the revised estimated within-lab SD listed in the enclosed Information sheet to evaluate their lab's quality control when using VITROS Performance Verifier I and II products. When available on the Ortho website, https://www.orthoclinicaldiagnostics.com, customers can obtain the updated VITROS Performance Verifier Control Assay Sheets and can use the revised ROM to evaluate the labs VITROS AMYL Slide quality control when using VITROS Performance Verifier I and II products. The firm explained that a review of previously reported patient results is NOT recommended. The customers were asked to Complete the Confirmation of Receipt form no later than May 19, 2020 and to forward the notification if the product was distributed outside of their facility.
- Quantity in commerce
- 73,071
- Distribution
- Worldwide distribution - US Nationwide including in the statesof Maine, Massachusetts, Vermont, Connecticut, New Jersey, New York, Pennsylvania, Washington DC, Maryland, Virginia, West Virginia, North Carolina, South Carolina, Georgia, Florida, Alabama, Tennessee, Missouri, Mississippi, Kentucky, Ohio, Indiana, Michigan, Iowa, Wisconsin, Minnesota, South Dakota, North Dakota, Montana, Illinois, Kansas, Nebraska, Louisiana, Arkansas, Oklahoma, Texas, Colorado, Idaho, Utah, Arizona, New Mexico, California, Hawaii, Oregon , Washington, and Alaska
- Product codes and lots
- Lot #s 600203564949, 600203564950, 600203564975, 600203564976, 600203564977, 600203564978, 600203566275, 600203566276, 600203566277, 600203566537, 600203566538, 600203566620, 600203566621, 600203566622, 600203566698, 600203566699, 600203566700, 600203567485, 600203567486, 600203567546, 600203567547, 600303570416, 600303570417, 600303570476, 600303570495, 600303570496, 600303571564, 600303571565, 600303576365, 600303576366, 600303578980, 600303578981, 600303578982, 600303578984, 600303578985, 600303579304, 600303579305, 600303579338, 600303579339, 600303579545, 600303579546, 600303579570, 600303579571, 600503581579, 620503581598, 600503581599, 600503581600, 600503581626, 600503582485, 600503582661, 600503582711, 600503582712, 600503582713, 600503583174, 600503583175, 600503584118, 600503584168, 600503584169, 600503589810, 600503589811, 600503589812, 600603591744, 600603591745, 600603591746, 600603595598, 600603595599, 600603595644, 600603595645, 600603595655, 600603595656, 600603597378, 600603597379, 600603599568, 600603599572, 600603599630, 600803637422, 600803637423, 600803637424, 600803638469, 600803638470
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2364-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2365-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2366-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2367-2020 | Class II | Terminated | Voluntary: Firm initiated |