Pump Tube Set for MMS Libra and Solar Systems -
LABORIE MEDICAL TECHNOLOGIES, CANADA ULC began a recall of Pump Tube Set for MMS Libra and Solar Systems - Product Usage: Intended for use during water perfusion urodynamic procedures to study the function of the... on . FDA classified it as Class II and gives the reason as "This field correction is being initiated to address potentially damaged seals in the packaging of some the Laborie Pump Tube Sets. Finished goods inventory was detected with damage to their seals in the form of breaches...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2535-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Laborie Medical Technologies, Canada ULC
- Product
- Pump Tube Set for MMS Libra and Solar Systems - Product Usage: Intended for use during water perfusion urodynamic procedures to study the function of the bladder and urethra during filling and voiding, by monitoring pressure and pressure changes.
- Product code
- FEZ Catheter And Tube, Suprapubic
- Reason for recall
- This field correction is being initiated to address potentially damaged seals in the packaging of some the Laborie Pump Tube Sets. Finished goods inventory was detected with damage to their seals in the form of breaches or folds. Based on our investigation the issue appears to be limited to a single lot of part number 902101226 pump tubing sets.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- This field correction is being initiated to address potentially damaged seals in the packaging of some the Laborie Pump Tube Sets. Finished goods inventory was detected with damage to their seals in the form of breaches or folds. Based on our investigation the issue appears to be limited to a single lot of part number 902101226 pump tubing sets.
- Quantity in commerce
- 480 units
- Distribution
- US Nationwide distribution including in the states of CO, FL, IN, MI, NJ, NY, OH, OR, PA and TX.
- Product codes and lots
- Model Number: REF 902101226, LOT 15883/1
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2535-2020 | Class II | Terminated | Voluntary: Firm initiated |