Z-2535-2020Class II

Pump Tube Set for MMS Libra and Solar Systems -

Laborie Medical Technologies, Canada ULC / initiated 2020-04-22 / Terminated / root cause: packaging

LABORIE MEDICAL TECHNOLOGIES, CANADA ULC began a recall of Pump Tube Set for MMS Libra and Solar Systems - Product Usage: Intended for use during water perfusion urodynamic procedures to study the function of the... on . FDA classified it as Class II and gives the reason as "This field correction is being initiated to address potentially damaged seals in the packaging of some the Laborie Pump Tube Sets. Finished goods inventory was detected with damage to their seals in the form of breaches...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2535-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Laborie Medical Technologies, Canada ULC
Product
Pump Tube Set for MMS Libra and Solar Systems - Product Usage: Intended for use during water perfusion urodynamic procedures to study the function of the bladder and urethra during filling and voiding, by monitoring pressure and pressure changes.
Product code
FEZ Catheter And Tube, Suprapubic
Reason for recall
This field correction is being initiated to address potentially damaged seals in the packaging of some the Laborie Pump Tube Sets. Finished goods inventory was detected with damage to their seals in the form of breaches or folds. Based on our investigation the issue appears to be limited to a single lot of part number 902101226 pump tubing sets.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
This field correction is being initiated to address potentially damaged seals in the packaging of some the Laborie Pump Tube Sets. Finished goods inventory was detected with damage to their seals in the form of breaches or folds. Based on our investigation the issue appears to be limited to a single lot of part number 902101226 pump tubing sets.
Quantity in commerce
480 units
Distribution
US Nationwide distribution including in the states of CO, FL, IN, MI, NJ, NY, OH, OR, PA and TX.
Product codes and lots
Model Number: REF 902101226, LOT 15883/1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2535-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2535-2020Class IITerminatedVoluntary: Firm initiated