Z-2407-2020Class III

Utak NR Trace Elements Serum Control, 5 ml, REF 66816. The responsible firm on the label is UTAK...

Utak Laboratories Inc / initiated 2020-03-20 / Terminated / root cause: labeling and UDI

Utak Laboratories Inc began a recall of Utak NR Trace Elements Serum Control, 5 ml, REF 66816. The responsible firm on the label is UTAK LABORATORIES, INC., Valencia, CA. Control for in vitro... on . FDA classified it as Class III and gives the reason as "The product has an incorrect reference value for cadmium and an incorrect expected range". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2407-2020
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Utak Laboratories Inc
Product
Utak NR Trace Elements Serum Control, 5 ml, REF 66816. The responsible firm on the label is UTAK LABORATORIES, INC., Valencia, CA. Control for in vitro diagnostics.
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
The product has an incorrect reference value for cadmium and an incorrect expected range.
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
The recalling firm issued a recall letter dated 3/19/2020 via email on 3/20/2020 requesting customers to destroy the incorrect IFU and replace it with the revised IFU.
Quantity in commerce
1,395 vials
Distribution
Worldwide Distribution. US nationwide, Canada, Australia, France, Germany, Israel, Italy, New Zealand, Philippines, and United Kingdom.
Product codes and lots
Lot C2540

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2407-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2407-2020Class IIITerminatedVoluntary: Firm initiated