Z-2401-2020Class II
Alinity C, Processing Module. Chemistry analyzer for in-vitro diagnostics
Abbott Medical / initiated 2020-05-12 / Terminated / root cause: design
Abbott Gmbh & Co. KG began a recall of Alinity C, Processing Module. Chemistry analyzer for in-vitro diagnostics on . FDA classified it as Class II and gives the reason as "There is a potential to generate incorrect patient results for the following assays: Alkaline Phosphatase, Amylase, Creatine Kinase, and Gamma Glutamyl Transferase". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2401-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Alinity C, Processing Module. Chemistry analyzer for in-vitro diagnostics.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- There is a potential to generate incorrect patient results for the following assays: Alkaline Phosphatase, Amylase, Creatine Kinase, and Gamma Glutamyl Transferase.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Abbott sent a notification dated May 12, 2020 titled, "Product Correction Urgent Immediate Action Required," to customers. This Product Correction letter is to inform customers of an issue with the Alinity c system Optics Assembly (part number 3501627801). The firm has identified the potential for a nonlinear optics response due to an issue with the Alinity c optics. The issue is with the diffraction grating within the optics assembly and can result in a lower than expected absorbance value. The root cause of the issue is unknown and currently under investigation. The issue can be identified using an Optics diagnostic procedure. Instruments that pass this Optics diagnostic procedure do not exhibit and will not develop the nonlinear optics response. Abbott has tested multiple representative assays with different calibration types and wavelengths to understand the impact of the nonlinear optics response on the firm's clinical chemistry menu. The studies compared sample results obtained between an Alinity c with a confirmed normal optics to the results obtained from an Alinity c with confirmed defective optics. The studies show that four assays are impacted by the nonlinear optics response: Alkaline Phosphatase, Amylase, Creatine Kinase, and GammaGlutamyl Transferase. The impact on the test result is a false decrease in results. An Appendix is attached to the notification regarding the assay testing information. There is a potential to generate incorrect patient results for the following assays on the Alinity c system: Alkaline Phosphatase, Amylase, Creatine Kinase, and GammaGlutamyl Tansferase. Abbott recommends discontinuing testing on the impacted assays (Alkaline Phosphatase, Amylase, Creatine Kinase, and GammaGlutamyl Transferase) until an assessment of the optics can be completed. A firm representative will contact the customer to schedule the assessment to determine the potential impact to the optics assembly. If customers have forwarded t
- Quantity in commerce
- 421 serial numbers
- Distribution
- U.S. States: AR, CA, CO, FL, GA, KS, KY, LA, MD, MI, MN, MO, MS, NC, NJ, NY, OK, SC, SD, TN, TX, UT, VA, WI, PR Worldwide distribution: ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BERMUDA, BRAZIL, CANADA, CHILE, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, INDONESIA, ISRAEL, ITALY, JAMAICA, JORDAN, KENYA, MALAYSAI, MEXICO, NEW CALEDONIA, NORWAY, OMAN, PAKISTAN, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, SAUDIA ARABIA, SERBIA, SINGAPORE, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, TURKEY, UAE, UK, URUGUAY, and VIETNAM.
- Product codes and lots
- List Number: 03R67-01: All serial numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2401-2020 | Class II | Terminated | Voluntary: Firm initiated |