Z-0518-2021Class II

MiniMed 640G, insulin pump, Ref/Model # MMT-1711K, MMT-1752K, MMT-1752, Rx Only -

Medtronic, Inc. / initiated 2020-04-30 / Open, Classified / root cause: manufacturing process

Medtronic Inc. began a recall of MiniMed 640G, insulin pump, Ref/Model # MMT-1711K, MMT-1752K, MMT-1752, Rx Only - Product Usage: indicated for the continuous delivery of insulin, at set and... on . FDA classified it as Class II and gives the reason as "Missing solder battery connection, which could interrupt power of insulin pump". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0518-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
MiniMed 640G, insulin pump, Ref/Model # MMT-1711K, MMT-1752K, MMT-1752, Rx Only - Product Usage: indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin.
Product code
N/A
Reason for recall
Missing solder battery connection, which could interrupt power of insulin pump.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On April 30, 2020, the firm mailed a "Urgent Medical Device Recall" letter to patients. The firm also sent a letter to Health Care Professionals, (HCP) "Urgent Medical Device Recall" that they read this notice and confirm receipt that the customer read and understood the instructions provided in this letter. The letter notified customers that the insulin pumps had an issue on the battery connection that did not meet firm's quality specifications. If power is interrupted on a patients insulin pump, then insulin is suspended, which may lead to hyperglycemia (high blood sugar). There have been no reported complaints from customers as a result of this issue. The firm identified the pump models, serial numbers of the affected devices. In addition to informing patients about the recall, the letter asked patients to do the following: (1) Record current pump settings. Follow the System User Guide to set up new pump that is enclosed in the box. (2) After successfully programming the replacement pump, start using replacement pump. (3) Return affected pump using the included shipping materials. Contact 24-Hour Technical Support at 1-877-585-0166, the instructions provided in this letter. The firm sent a letter to Health Care Professionals, (HCP) "Urgent Medical Device Recall" that they read this notice and confirm receipt that the customer read and understood lead to hyperglycemia (high blood sugar). There have been no reported complaints from customers as a result of this issue. The firm identified the pump models, serial numbers of the affected devices.
Quantity in commerce
2
Distribution
Worldwide distribution - U.S. Nationwide Distribution in the state of MD and the countries of Canada, Netherlands, Norway. ***Model 630G distributed in Canada and Model 670G distributed in US ***MODEL640G Distributed OUS ONLY*** OUS: Australia, Canada, Netherlands and Norway.
Product codes and lots
All serial numbers expiration date- 3 years from date of manufacture: NG2095284H, NG2095319H

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0518-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0516-2021Class IIOngoingVoluntary: Firm initiated
Z-0517-2021Class IIOngoingVoluntary: Firm initiated
Z-0518-2021Class IIOngoingVoluntary: Firm initiated