Siemens IMMULITE 2000 XPi Immunoassay System - In vitro Chemistry Analyzer SMN #10373214 -
Siemens Healthcare Diagnostics Inc. began a recall of Siemens IMMULITE 2000 XPi Immunoassay System - In vitro Chemistry Analyzer SMN #10373214 - Product Usage: are intended for professional use in a laboratory... on . FDA classified it as Class II and gives the reason as "Flexible tubing connected to the liquid waste bottle can crack during routine customer maintenance potentially causing liquid waste can leak onto the floor, creating a slip and fall hazard". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2470-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Siemens IMMULITE 2000 XPi Immunoassay System - In vitro Chemistry Analyzer SMN #10373214 - Product Usage: are intended for professional use in a laboratory environment only.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Flexible tubing connected to the liquid waste bottle can crack during routine customer maintenance potentially causing liquid waste can leak onto the floor, creating a slip and fall hazard
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Siemens issued an Urgent Medical Device Correction (UMDC IMI20-01.A.US) to US customers via FedEx on May 15, 2020. Urgent Field Safety Notice (UFSN IMI20-01.A.OUS) was electronically issued to all OUS Siemens sites with affected customers, for distribution, as well as regional reporting to their respective local competent authority on May 15, 2020. The field action letters explain the issue, health risk and and inform customers of actions to take: Siemens Customer Service Engineers shall be visiting customer sites to replace Waste Bottle Flex Tubing Assemblies. Customers are requested to return the Effectiveness Check Form, in the UMDC/UFSN via fax or email.
- Quantity in commerce
- 2925 units
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of AE, AO, AR, AT, AU, AZ, BA, BD, BE, BG, BH, BO, BR, BS, BY, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GP, GR, HK, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JO, JP, KR, KW, KZ, LB, LT, LU, LV, LY, ME, MK, MM, MQ, MT, MU, MX, MY, NL, NO, NZ, PE, PH, PK, PL, PT, RO, RS, RU, SA, SE, SG, SI, SK, SY, TH, TN, TR, TT, TW, UA, UY, VE VN, ZA.
- Product codes and lots
- Serial numbers A001 through M5295
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2469-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2470-2020 | Class II | Terminated | Voluntary: Firm initiated |