Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2258-2020Class II

System, Tomography Computed Emmission

GE Healthcare / initiated 2018-11-30 / Open, Classified / root cause: software

GE Healthcare, LLC began a recall of System, Tomography Computed Emmission on . FDA classified it as Class II and gives the reason as "GE Healthcare has identified that a small number of Nuclear Medicine systems are operating without currently updated software. In a recent event, a patient s finger was injured during the Unload process while the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-2258-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
System, Tomography Computed Emmission
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
GE Healthcare has identified that a small number of Nuclear Medicine systems are operating without currently updated software. In a recent event, a patient s finger was injured during the Unload process while the automatic Unload feature was being used; the software on the system was not current.
Root cause tag
software
FDA root cause
Other
Action
GE Healthcare identified that a small number of Nuclear Medicine systems are operating without currently updated software.
Quantity in commerce
526 total units
Distribution
Domestic Distribution: AL, AR, AZ, CA, CO, CT, DC, DE, FL,GA, IA, IL, IN, KY, LA,MA, MD, ME, MI, MN, MO, MS, NC, NE, NM, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT,VA, WI and WV. Foreign Distribution: Argentina, Australia, Belgium, Brazil, Canada, Colombia, Denmark, France, Germany, Ghana, India, Israel, Italy, Japan, Korea, Malaisya, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Sweden, Taiwan, Thailand, Turkey and United Kingdom.
Product codes and lots
Model for Ventry: ASM001605, H3000YT, H3000YW, H3000YY, H3000ZW  Serial Numbers: 11075,11430,11137,11154,11138,11512,11322,11093,11301,11286,11294,11021,11078,11125,11235,11387, 11337, 11288,11396,11147,11039,11427,11328,11471,11251,11424,11448,11421,11450,11203,11468,11363,19016,11069,11127,11284,11124,11229,11057,11143,11515,11190,11001,11027,11297,11306,11061,11151,11161,11441,11248,11067,11247, 11435, 11077, 11513, 11497,  11206,11211,11347,19069,11085,11392,11169,11222,11325,19098,19082,11398,11420,11378,11166,11008,11470,19070,19086,19027 19044,19083,19065,19062,11464,11280,11469,11376,11310,11119,11501,19048,11076,11086,11031,11366,11214,11383,19009,11345, 11358,11183,11422,11004,11307,11074,11503,11276,11054,11270,11189,11142,11244,11060,11145,11260,11092,11431,11038,11170,19109,19015,19040,11087,11462,11465,11117,19085,11179,11414,11150,19012,19108,11365,11255,11176,11089,11374,11017,11279,11359, 11324,11146,11312,11320,19030,11272,11360,11371,11385,11467,11231,11511,11505,11059,11045,11230,11353,11409,11349,11148,11315,11410,11044,11461,19057,11228,11443,11123,11237,11082,11090,11303,11438,11287,11454,11491,11065,11084,11080,11212,11269, 11489,11517,11043,11299,11300,11213,11278,11417,11053,11070,11122,11025,11098,11372,11391,11393,11397,11389,11232,11234, 11063,11180,19002,11018,11051,11096,11485,16001,16007,19008,19097,11095,11210,11490,11030,11033,11309,11022,11403,11330, 11340,11162,11442,11014,11407,11514,11277,11402,11066,11167,11164,11226,11426,11308,11445,11208,11282,11404,11304,11185, 11181,11227,11264,11267,11289,11415,11311,11333,11331,11026,11412,11023,11290,11346,11405,11482,11221,11224,11369,11002, 11003,11118,11444,11449,11344,11029,11439,11182,11436,11356,11423,11478,11494,11215,11246,19072,11058,19080,11105,11475, 11472,11486,11032,11201,11113,11474,11296,16020,11091,11357,11121,11159,11252,11463,11498,11406,11487,11271,11437,11321, 11009,11216,11377,11132,11458,11133,11367,11516,19026,11317,11064,11195,11072,11305,11452,11219,11316,11373,11492,19005, 11062,11382,11245,11314,11103,11394,11400,11429,11432,11433,11323,11342,11007,11205,11202,11281,11129,11225,11010,19025, 11097,11171,11350,11368,11049,11268,11131,11481,11186,11476,00000000VEN002, 0000E3GT09163A, 0000E3GT08038A, 00000000VEN003.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2258-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2258-2020Class IIOngoingVoluntary: Firm initiated
Z-2259-2020Class IIOngoingVoluntary: Firm initiated