Z-2432-2020Class II
RP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11416778
Siemens Healthcare Diagnostics Inc. / initiated 2020-05-27 / Terminated / root cause: software
Siemens Healthcare Diagnostics Inc began a recall of RP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11416778 on . FDA classified it as Class II and gives the reason as "The scanner is not confirming the integrity of read barcode data using the check-digit when reading Code 39 barcodes". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2432-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- RP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11416778
- Product code
- CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
- Reason for recall
- The scanner is not confirming the integrity of read barcode data using the check-digit when reading Code 39 barcodes.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On May 27, 2020, the firm notified customers via Urgent Field Safety Notice letters. Product was distributed to foreign customers only. Customers were instructed to complete and return the included customer response form. Customers whose systems are not configured to use code 39 with check-digit need take no further action. However, for customers whose systems are configured to use code 39 with check-digit, the customer notice provided two workaround options: 1) Use the integrated on-board barcode scanner, or 2) Re-configure the handheld barcode scanner following the instructions in the letter. If you have any questions, please contact your Siemens Customer Care Center or your local Siemens technical support representative
- Quantity in commerce
- 436
- Distribution
- Foreign distribution only.
- Product codes and lots
- All units
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2432-2020 | Class II | Terminated | Voluntary: Firm initiated |