Z-2259-2020Class II
System, Tomography Computed Emmission
GE Healthcare / initiated 2018-11-30 / Open, Classified / root cause: software
GE Healthcare, LLC began a recall of System, Tomography Computed Emmission on . FDA classified it as Class II and gives the reason as "GE Healthcare has identified that a small number of Nuclear Medicine systems are operating without currently updated software. In a recent event, a patient s finger was injured during the Unload process while the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2259-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Healthcare
- Product
- System, Tomography Computed Emmission
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- GE Healthcare has identified that a small number of Nuclear Medicine systems are operating without currently updated software. In a recent event, a patient s finger was injured during the Unload process while the automatic Unload feature was being used; the software on the system was not current.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- GE Healthcare identified that a small number of Nuclear Medicine systems are operating without currently updated software.
- Quantity in commerce
- 526 total units
- Distribution
- Domestic Distribution: AL, AR, AZ, CA, CO, CT, DC, DE, FL,GA, IA, IL, IN, KY, LA,MA, MD, ME, MI, MN, MO, MS, NC, NE, NM, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT,VA, WI and WV. Foreign Distribution: Argentina, Australia, Belgium, Brazil, Canada, Colombia, Denmark, France, Germany, Ghana, India, Israel, Italy, Japan, Korea, Malaisya, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Sweden, Taiwan, Thailand, Turkey and United Kingdom.
- Product codes and lots
- Model for Discovery NM 530c: 5248934, 5166674-01, H3000SA, H3001JA, H3001JD. Serial Numbers: 19016,0000000ME11041,19069,19098,19082,19070,19086,19027,19044,19083,19065,19062,19048,19009,19109,19015,19040,19085,19012, 19108, 19030, 19057,000000109961-1,19002,16001,16007,19008,19097,19072,19080,16020,19026,19005,19025,0000000000ME-4.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2258-2020 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2259-2020 | Class II | Ongoing | Voluntary: Firm initiated |