TactiCath Quartz Contact Force Ablation Catheter, Model Numbers: PN-004 065 (GTIN 07640157990033) and PN-004...
Abbott began a recall of TactiCath Quartz Contact Force Ablation Catheter, Model Numbers: PN-004 065 (GTIN 07640157990033) and PN-004 075 (GTIN 07640157990040) - Product Usage: is a... on . FDA classified it as Class II and gives the reason as "High rate of reported complaints for the loss of contact force information while using the TactiCath Quartz Contact Force Ablation Catheter". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2541-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- TactiCath Quartz Contact Force Ablation Catheter, Model Numbers: PN-004 065 (GTIN 07640157990033) and PN-004 075 (GTIN 07640157990040) - Product Usage: is a multi-electrode irrigated catheter with a deflectable tip designed to allow electrophysiological mapping of the heart and, when connected to an RF generator, to deliver radiofrequency current to the catheter tip electrode for the purpose of ablation.
- Product code
- OAE Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Fibrillation
- Reason for recall
- High rate of reported complaints for the loss of contact force information while using the TactiCath Quartz Contact Force Ablation Catheter
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm began notifying their consignees by letter on 05/22/2020. The notification reported the high rate of reported complaints for loss of contact force information while using the product, the risk to patient, and recommendations to avoid contact force failures.
- Quantity in commerce
- 84104 devices
- Distribution
- Worldwide distribution.
- Product codes and lots
- All batch numbers with an expiration date of April 25, 2020
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2541-2020 | Class II | Terminated | Voluntary: Firm initiated |