Z-2522-2020Class II

uCT 760 System, Computed Tomography X-ray System

Shanghai United Imaging Healthcare Co., Ltd. / initiated 2020-05-08 / Terminated / root cause: software

Shanghai United Imaging Healthcare Co., Ltd. began a recall of uCT 760 System, Computed Tomography X-ray System on . FDA classified it as Class II and gives the reason as "Two issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and intermittent scout scanning interruption due to occasional angle...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2522-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Shanghai United Imaging Healthcare Co., Ltd.
Product
uCT 760 System, Computed Tomography X-ray System
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Two issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and intermittent scout scanning interruption due to occasional angle signal drift. This may result in the need for rescanning of the patient with the patient receiving an additional dose of radiation and possible use of additional contrast media.
Root cause tag
software
FDA root cause
Software design
Action
Customers were sent emails on 05/22/2020 notifying them of the failure modes and the potential safety issues. Customers were informed that If image artifacts occur, it may be necessary for the user to repeat the examination. The rescan of the patient will result in an additional dose and the possible need for additional contrast medium. Customers were also informed that if the scout scanning is unexpectedly interrupted, it may be necessary for the user to repeat the scout scanning, resulting in an additional dose. Customers observing image artifacts were instructed to reboot the system and continue the reconstruction, and, if image artifacts persist, to contact their United Imaging service support team or call the United Imaging Customer Contact Center at (855) 221-1552 (toll free) or (832) 699-2799 (direct). Similarly, if scout scanning is interrupted, customers were instructed to attempt repeating the scout and to contact the service support team if the problem continues. The email informs customers that United Imaging has released updated software R001.11.0.820958-Re-1150SP1+H1H2 to solve the intermittent issue and to eliminate angulation signal drift, which could eliminate the resulting image artifacts and scout scanning interruption, and to accomplish other improvements. Customers with further questions should contact the United Imaging Customer Contact Center at {855) 221-1552 {toll free) or {832) 699-2799 {direct).
Quantity in commerce
5 units
Distribution
Distributed to two customers. Both customers located in Houston, Texas.
Product codes and lots
Model uCT 760; Serial Numbers: 600229, 600230, 600232, 600271 with software version R001.11.0.729741-Re-1100, and serial number 600282 with software version R001.11.0.785930-Re-1150.