uCT 530 Computed Tomography X-Ray System

Shanghai United Imaging Healthcare Co., Ltd. / initiated 2020-05-08 / Open, Classified / root cause: software

Shanghai United Imaging Healthcare Co., Ltd. began a recall of uCT 530 Computed Tomography X-Ray System on . FDA classified it as Class II and gives the reason as "Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-2472-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Shanghai United Imaging Healthcare Co., Ltd.
Product
uCT 530 Computed Tomography X-Ray System
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.
Root cause tag
software
FDA root cause
Software design
Action
The firm sent letters to their customers by email on 5-22-2020. The letter informs customers of the two potential safety issues with the device, and notes that these problems may cause image artifacts and scanning workflow interruptions. Risks to patient health are noted to include the possible need to rescan the patient resulting in an additional radiation dose and the possible need for additional contrast medium. Customers were informed that United Imaging has released a field update for the uCT 530 to address the issues. If customers observe image artifacts, they are requested to reboot the system and continue the reconstruction. If the scout scanning is interrupted, customers are instructed to try repeating the scout scanning. If problems persist, customers should contact their United Imaging service support team or call the United Imaging Customer Contact Center at (855) 221-1552 (toll free) or (832) 699-2799 (direct).
Quantity in commerce
4 units
Distribution
Distributed to Texas.
Product codes and lots
Units with software version no. R001.11.0.729741-Re-1100 including the following: Serial number/GTIN: 300162 (GTIN: 6971576831012); 300163 (GTIN: 6971576831012);  300174 (GTIN: 6971576831012); and 300176 (GTIN: 6971576831012).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2472-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2472-2020Class IIOngoingVoluntary: Firm initiated