Ion Catheter Reprocessing Cover, IF1000 (Part Number:490114-02) UDI:00886874116845
Intuitive Surgical, Inc. began a recall of Ion Catheter Reprocessing Cover, IF1000 (Part Number:490114-02) UDI:00886874116845 on . FDA classified it as Class II and gives the reason as "One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2949-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- Ion Catheter Reprocessing Cover, IF1000 (Part Number:490114-02) UDI:00886874116845.
- Product code
- EOQ Bronchoscope (Flexible Or Rigid)
- Reason for recall
- One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or water during reprocessing; may cause failure of the ION Fully Articulating Catheter or ION System.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm notified their consignees regarding the "Urgent Medical Device Recall" notification via Fedex, and issued on 06/10/2020. The actions to be taken by the Customer/User are as follows: Locate the affected product at the site and return to the recalling firm via the standard RMA process. As part of the standard field action process, please take the following actions: 1. Inform necessary hospital personnel about this issue. 2. Inform necessary personnel when corrective action has been completed. 3. Complete the Acknowledgement Form and return to the recalling firm as instructed on the form. 4. Please retain a copy of letter and Acknowledgement Form for reference. Actions to be taken by the recalling firm: 1. A firm representative will make contact to support in replacing the affected product. 2. The firm will promptly overnight replacement product (PN 490118) after Customer Service is contacted by phone and receives the number of affected Catheter Reprocessing Covers (PN 490114-02) at the hospital. If further information is needed or support concerning the Medical Device Recall, contact your Sales Representative or contact Customer Service at the number listed below: Customer Support: 1-866-225-2071, Option 3 (7 AM to 5 PM EST) or [email protected].
- Quantity in commerce
- 67 units
- Distribution
- U.S.: California, Colorado, Florida, Illinois, Massachusetts, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Texas, and Virginia. No foreign consignees.
- Product codes and lots
- All lots.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2949-2020 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2950-2020 | Class II | Ongoing | Voluntary: Firm initiated |